Research Article

e-SCP- Protocol: An Expansion on SCP-ECG Protocol for Health Telemonitoring—Pilot Implementation

Table 3

New defined parameters stored in Section-1.

TagLengthValue (Parameter Data)

200length Second patient ID (Text characters)

2012 Patient Nationality (Binary)
This has the following format:
Byte Contents
1-2 binary: Nationality indication (ISO 3166.1) defined as:
0 Unspecified

202length Patient Address (Text characters)

203length Patient Phone Number (Text characters)

2041 Patient Religion (Binary)
This has the following format:
Byte Contents
1 Binary: set equal to 255
0Unspecified
1Atheist
2Baha’I
3Buddhism
4Christianity
5Confucianism
6Hinduism
7Islam
8Jainism
9Judaism
10Shinto
11Sikhism
12Daoism
13Zoroastrianism
14–30Reserved
31–255Manufacturer specific

205length Birth Place (Text characters)

2061 Patient Insurance (Binary)
This has the following format:
Byte Contents
1-2 binary: Nationality indication (ISO 3166.1) defined as:
0Unspecified

207length Memorial History (Free Text)
This field contains a text description of hereditary diseases.

2081 Blood Type (Binary)
This has the following format:
Byte Contents
1 binary: set equal to 255
0Unspecified
1A+
2A
3B+
4B
5AB+
6AB
7O+
8O
209length Profession (Free Text)
This field contains a text description of people profession.

2101 File access (Binary)
This has the following format:
Byte Contents
1 Binary: type of file access
0read/write
1read only
2Locked

211length Access restrictions (Binary)
This has the following format:
Byte Contents
1 Binary: type of access restrictions
0All file contents available to all medical specialties
1File contents available depended to medical specialties
2All file contents available to specific medical specialty
2 Binary: if byte 1 has value different from 0 then this field contains the medical specialty. There is no limit on the number of specialties.

212length SPO2 Machine ID Acquiring Device (Binary bytes and Text characters)

213length NiBP Machine ID Acquiring Device (Binary bytes and Text characters)

214length CO2 Machine ID Acquiring Device (Binary bytes and Text characters)

215length GPS Machine ID Device (Binary bytes and Text characters)

2161 Operational mode (Binary)
This has the following format:

Byte Contents
10Basic
1Emergency

* the numeric values and ** the description of the corresponding codes of countries in ISO 3166-1.