Research Article

Performance of an Ultrasonic Imaging System Based on a 45-MHz Linear PVDF Transducer Array: A Numerical Study

Table 1

Material constants used in the numerical simulations [2023]. A PVDF film with Young’s modulus and electromechanical coupling factor of 4.0 GPa and 14%, respectively, was chosen. Properties of PVDF films vary depending on the processing technique, temperature, frequency, and age [2426]. Attenuation in PVDF, epoxy, copper, and human skin are based on data from [23, 2729].

PVDFEpoxyCu AuSkin

Density (103 kg/m3)1.78 1.208.9619.31.00
Young’s modulus (109 Pa) 4.0 3.0 120 79
Poisson’s ratio (−) 0.20 0.340.310.46
Longitudinal wave speed (103 m/s) 1.6 2.0 4.34.41.54
Transverse wave speed (103 m/s) 1.0 1.0 2.3 1.2 0
Attenuation (dB/(cm MHz)) 10 1.20.03 2

Piezoelectric const. (C/m2) 0.004
Piezoelectric const. (C/m2) −0.096
Piezoelectric const. (C/m2) 0
Permittivity (10−12 F/m) 106