An Ultrasound Imaging-Guided Robotic HIFU Ablation Experimental System and Accuracy Evaluations

Table 1

The distance error of three peak points (of the calibration template) (unit: mm).

Number of pointCoordinates of the target pointNumber of imageCoordinates of the guided pinpointDistance error
xyzxyz

Case A: 3 cm
126.78.9−46.5I26.68.6−46.80.44
II27.48.4−46.40.87
III27.18.4−46.70.70

26.2−5.4−45.2I5.3−4.5−44.91.31
II6.0−5.2−44.80.49
III6.3−5.0−44.60.73

31.8−19.1−61.5I2.0−19.3−62.00.57
II1.9−19.6−61.90.65
III1.8−19.4−61.60.31

Max error: 1.31; average error: 0.67; standard deviation 0.27

Case B: 7 cm
138.927.9−89.6I39.228.1−90.50.97
II39.128.0−90.71.12
III39.627.8−90.51.14

218.314.0−87.7I17.513.8−88.51.15
II18.013.6−88.10.64
III18.113.9−88.20.55

313.7−0.1−104.5I14.1−0.5−105.41.06
II14.00.0−105.91.43
III14.2−0.2−105.51.12

Max error: 1.43; average error 1.02; standard deviation: 0.26

Case C: 12 cm
154.146.8−136.3I55.446.7−136.21.31
II53.547.5−137.61.59
III53.947.3−137.31.14

233.232.9−134.6I34.532.8−134.21.36
II32.733.4−135.30.99
III32.832.9−135.40.89

328.919.0−151.6I29.819.4−151.41.00
II28.620.1−152.11.24
III28.019.2−152.91.59

Max error: 1.59; average error: 1.24; standard deviation: 0.24