Juana-Mariel Dávila-Vilchis, Juan C. Ávila-Vilchis, Adriana H. Vilchis-González, LAZ-Avilés, "Design Criteria of Soft Exogloves for Hand Rehabilitation-Assistance Tasks", Applied Bionics and Biomechanics, vol. 2020, Article ID 2724783, 19 pages, 2020. https://doi.org/10.1155/2020/2724783

Design Criteria of Soft Exogloves for Hand Rehabilitation-Assistance Tasks

Academic Editor: Simo Saarakkala
Received24 Feb 2020
Revised03 Jul 2020
Accepted07 Jul 2020
Published01 Aug 2020

Abstract

This paper establishes design criteria for soft exogloves (SEG) to be used as rehabilitation or assistance devices. This research consists in identifying, selecting, and grouping SEG features based on the analysis of 91 systems that have been proposed during the last decade. Thus, function, mobility, and usability criteria are defined and explicitly discussed to highlight SEG design guidelines. Additionally, this study provides a detailed description of each system that was analysed including application, functional task, palm design, actuation type, assistance mode, degrees of freedom (DOF), target fingers, motions, material, weight, force, pressure (only for fluids), control strategy, and assessment. Such characteristics have been reported according to specific design methodologies and operating principles. Technological trends are contemplated in this contribution with emphasis on SEG design opportunity areas. In this review, suggestions, limitations, and implications are also discussed in order to enhance future SEG developments aimed at stroke survivors or people with hand disabilities.

1. Introduction

Hand and finger motions are imperative for grasping and manipulation tasks. Nonetheless, people who have suffered from cerebral palsy (CP), stroke, or spinal cord injury (SCI) have great difficulty in accomplishing these activities of daily living (ADL) by themselves. A person with any of these pathologies could present clenched fist, spasticity, uncoordinated motions, loss of strength, or diminished dexterity. These are consequences of a neuronal impairment that is responsible for controlling motricity, muscle endurance, and tonicity [1]. Worldwide, more than 15 million people are affected each year [2], and only 11.6% of the stroke survivors are able to recover dexterity [3]. Patients with these disabilities can, freely, flex their hand muscles but show abnormal resistance when extending them [4], requiring physical rehabilitation or assistance.

Other hand motor deficits are caused by ageing or hand deformities such as rheumatoid arthritis or osteoarthritis, because cartilage weakens, muscle mass decreases, and joint stiffness increases [5]. More than 50 million elderly people have difficulties to achieve accurate gripping and pinching forces, and their range of motion (ROM) is limited as well as their work area [6].

Therefore, people with hand disabilities can initiate a prompt rehabilitation protocol in order to start recovering motor skills, stop joint stiffness, and increase their independence and self-esteem [7]. Physical and occupational therapies are the most common treatments to recover patients’ movements, for example, adduction-abduction or flexion-extension of finger, wrist, or elbow joints. However, these routines can be exhausting, time-consuming, and, relatively, costly since patients require the assistance of a therapist whose availability is uncertain [8].

Normally, rehabilitation programs are customized for each patient due to their impairment, age, and anthropometric dimensions. Moreover, these robot procedures are classified into three main assistance levels: passive assisted mode (PAM), active assisted mode (AAM), and active resistive mode (ARM) depending on the recovery status of patients and support of a robot [9].

Literature has reported that rehabilitation protocols can be executed by robots or soft wearable devices which have emerged as a therapy tool with safe human interactions, low weight, and affordable systems [10]. Particularly, SEG have become an alternative approach in the effort to overcome hand dysfunctions and assist patients with handling tasks. SEG have the ability to combine conventional therapy with wearable systems to mimic the natural movement of fingers in order to increase their mobility, preventing spasticity and joint stiffness [11].

SEG have mainly evolved in terms of their design, fabrication, and control [12]. Pioneering designs started using sport gloves incorporating a control system [13, 14]. Then, SEG proposals explored synthetic leather [15], rubber [16, 17], and fabrics [18, 19] to provide flexible human-robotic interactions as in the case of bike gloves [20]. Elastomers have become the primary option to empower flexibility and lightness [21]. Moreover, instead of closed palm designs (CPD) where the whole hand is covered with the glove, open palm designs (OPD) with bare hands use elastomers trying to behave as a natural extension of the human hand to compete with skin properties in order to achieve a suitable contact with objects [22, 23]. Other assistance SEG have been developed for material handling in hazardous environments, support in heavy-lifting tasks [24, 25], or extravehicular tasks in space [26].

Mostly, SEG systems have been driven by electrical energy or fluid (pneumatic or hydraulic) pressurization. Regarding electrical power supplies, tendon-driven systems employ linear actuators to push and pull cables embedded in Teflon tubes [27]. Pneumatic actuation includes fiber-reinforced elastomer actuators (FREAs), inflatable chambers, or pneumatic artificial muscles, commonly known as McKibben muscles [28].

People with hand dysfunctions demand for reliable SEG to improve their quality of life. Nevertheless, the lack of affordable and accessible SEG for hand impairment patients with low-cost manufacturing processes is still a significant challenge. Therefore, this paper has reviewed the progress in the field of SEG for neuromuscular rehabilitation and assistance to overcome hand motor dysfunctions.

The main contribution of this paper is the identification and classification of 13 design criteria to provide a set of guidelines for SEG developments based on an extensive review of the state of the art and of the technique from the last decade. Moreover, a detailed description of 91 SEG systems is provided along with implications, limitations, and suggestions for future developments.

This paper is organized as follows. Section 2 presents, classifies, and discusses the criteria that are proposed for SEG design based on reported devices and specific literature. Section 3 reports SEG’s development guidelines together with the characteristics of the 91 reviewed devices. Section 4 provides a discussion concerning significant aspects (limitations, implications, and suggestions) to be taken into account for future developments of SEG systems. Conclusions are at the end of this document in Section 5.

2. SEG Design Criteria

Hand mobility characterization in SEG designs has turned out to be a challenge since hand anatomy is one of the most complex kinematics parts of the human body with 20 DOF for the whole of the fingers: one for abduction-adduction in every finger (thumb included); 12 for flexion-extension for index, middle, ring, and pinkie fingers; and three for thumb including opponent motion [29].

In this paper, 2 function criteria, 6 mobility criteria, and 5 usability criteria are proposed in order to enhance SEG designs and enable fast developments. These design criteria are based on the aspects that have been identified from the 91 SEG systems reported in this article and on the soft wearable device’s methodology established in [28]. Figure 1 illustrates the proposed criterion classification.

Moreover, biological inspiration has come to the fore in SEG design to emulate an animal’s motion looking for stability [30] or optimal grasping tasks [31]. According to [8], SEG should weigh less than 500 g, provide easy and comfortable donn-doff, and achieve 10 open-close finger cycles per minute for effective actuation. Regarding SEG mechanical design, authors in [29] suggest taking into account the number of joints and working DOF, the type of actuators, and the application. Other attributes in SEG design should adopt the characteristics of a rehabilitation device which include mode of intervention (unilateral or bilateral), number of DOF, target portion (distal, proximal, or quantity), and motion guidance (passive or active), among others [32].

Based on reported literature, the next paragraphs discuss the criteria presented in Figure 1.

2.1. SEG Function Criteria

SEG are classified into rehabilitative or assistive devices depending on their purpose [14]. SEG systems must be able to execute physical therapy and manipulation tasks to offer efficient and competitive devices for those with hand disabilities. Then, rehabilitation and assistance criteria must consider the aspects discussed in the respective paragraphs.

2.1.1. Rehabilitation Criterion

Rehabilitation SEG are designed to help the patient regain strength, dexterity, and coordination to recover hand functionality and range of motion (ROM) [33]. These SEG are focused on performing specific fist motions such as full, hook, straight, and tabletop [18] or open-close to improve grasping tasks [34].

Thumb, index, and middle finger flexion-extension is needed for strong grasping [31, 35]. Supplementary motions such as adduction-abduction are required to grasp and release objects in a more natural way [36]. Furthermore, flexion at the interphalangeal (IP) and metacarpophalangeal (MCP) joints with rotation at the carpometacarpal (CMC) joint is necessary to reproduce thumb opposition [37]. Other SEG are able to perform wrist flexion [38], wrist radial-ulnar deviation [16], or forearm pronation-supination motion [39].

SEG rehabilitation routines can include virtual reality in order to analyse the effects of brain stimulation when executing specific tasks [40]. Patients are immersed in a game environment where they achieve manipulation tasks such as squeezing oranges, catching butterflies, or grabbing objects [39]. Other SEG rely on neuroimaging techniques [41] or provide feedback to assess a patient’s conditions and monitor their progress [42]. Nevertheless, it is not enough to train the brain and do physical therapy; a successful rehabilitation process depends on the patient’s response and their own capabilities [43].

Depending on each rehabilitation protocol, the required time to use a soft exoglove varies. For instance, 60 minutes per day is recommended by [44]. Pilot tests performed by [45] suggested rehabilitation sessions from 30 to 40 minutes 5 days a week. Authors in [38] recommend 45 minutes but no more than 90 minutes per day to avoid SEG strain deformations. Authors in [46] suggest 180 minutes per week, while authors in [39] determine that 30 minutes per day over the course of 20 sessions is necessary for a positive sizable impact on the impaired hand. Furthermore, to achieve a successful rehabilitation program, patients should combine 30 minutes of SEG training with 30 minutes on occupational therapy per day [47].

2.1.2. Assistance Criterion

Eating, dressing, and writing are everyday actions that are done unconsciously. Nevertheless, those tasks turn out to be a tough challenge for people with hand dysfunctions. Normally, patients depend on their family or on a therapist to assist them [48]. Hence, assistive SEG are intended to help patients to achieve manipulation tasks despite their restricted ROM, to interact with their surroundings, and to execute ADL by themselves. These systems are recommended when rehabilitative SEG are not enough to overcome patient stiffness [49].

SEG for assistance tasks are designed to perform three integral functions of the human hand: (i) finger mobilization, (ii) holding (grasping and gripping) with high precision and strength, and (iii) manipulation for positioning and releasing objects [8]. Assistive SEG should execute grasping, holding-lifting, and releasing motions as continuous actions to achieve a complete manipulation [50]. To achieve stable grasping, thumb, index, and middle fingers must be included on SEG systems [35]. According to [51], soft exoglove devices should provide 8 N of grasping force to manipulate an object with a mass of 1.5 kg.

2.2. SEG Mobility Criteria

From a functional perspective, authors in [52] propose that weight, size, and power consumption can define an efficient soft exoglove that fits the anatomical ROM of the human hand. The mass of the whole system should not exceed 3 kg to be considered as an assistive device [50]. These characteristics are included in criteria 3 to 8 (see Figure 1): actuation, materials, guidance mode, manufacture, operation and control, and assessment that are discussed as follows.

2.2.1. Actuation Criterion

As aforementioned, tendon-driven actuators use wires to emulate human tendon functions as flexion-extension motion. This type of actuation can include Bowden cable transmissions to separate the control unit from the end effector and reduce weight [53]. Also, artificial muscle wires have been proposed to avoid friction [54], and shape memory alloys (SMA) have been employed due to their elasticity [55] and high force-weight ratio [26].

On the other hand, pneumatic actuators could be embedded into inflatable air bladders [16] and into a double layer sheet with curved rubber muscles [15] or made of flexible electrostatic discharge plastic sheet materials [1, 56]. The McKibben muscles represent an affordable choice [57] and have the ability to constrain any radial expansion during pressurization [58]. Hydraulic actuators offer high load capacity [11].

A new trend is hybrid actuation which fits hand motion shape using soft pneumatic actuators and tendon-driven operation [7], providing customization based on rigid frames and soft muscles [48]. Table 1 reports the advantages and disadvantages of different SEG actuations.


ActuationTypesAdvantagesDisadvantages

Electrical(i) Muscle wires
(ii) Tendon-driven
(iii) Shape memory alloys
(i) Cable paths reduce friction
(ii) Provides continuous force
(iii) Stores energy
(iv) Commercial availability
(i) Complex transmissions
(ii) Continuous hours of operation are restricted
(iii) Nonlinearity of the system makes control difficult

Pneumatic(i) FREA
(ii) Inflatable chambers
(iii) Pneumatic artificial muscles
(i) Allows multiple DOF
(ii) Supports their structural shape
(iii) Allows adaptability
(iv) Lightweight
(i) Requires compressed air
(ii) Requires a reservoir
(iii) Inaccurate forces
(iv) Problems with leaks
(v) Portability is restricted

Hydraulic(i) Fluid chambers(i) High load capacity and power supply
(ii) Low cost
(iii) Allows multiple DOF
(i) Heavy systems
(ii) Problems with leaks
(iii) Portability is restricted
(iv) Requires a reservoir and a pump

When using a soft glove, patient safety must be guaranteed. Thus, all SEG must include different safety strategies and levels in their design. For example, on cable actuation, mechanical stops, torque, or tension limiters have been implemented [59]. Regarding pneumatic actuation, solenoid and exhaust valves are employed along with pressure regulators to control air flow or avoid air returns [41]. Quasistatic, dynamic, and material failures are discussed in [60], where measures that can be considered in order to avoid unsafe situations for soft robots are provided.

Other safety levels have been applied to the electrical configuration such as emergency stops, watch dogs, or physical decoupling of power interfaces from logic ones by electromagnetic couplings [51]. In addition, by using closed-loop control (CLC) schemes, sensing errors are minimized and operation in a stable regime is ensured to avoid hyperextension at the wrist or overflexed fingers, for instance [20]. At the programming level, haptic feedback is also included to prevent accidents [61].

More specialized safety strategies related to robots can be considered, such as safety standards or means to guarantee system dependability [62] as fault prevention, fault removal, fault forecasting, and fault tolerance [63]. Being safety a priority aspect, it constitutes a current research area by itself and must be taken into account in the development of SEG systems. Concerning rehabilitation robots, ISO-IEC 80601-2-78 must be taken into account. Many specialized documents are recommended for readers interested in this topic and for researchers and engineers working in SEG design (see, for instance, [6466]).

Additionally, relevant features for actuators have been identified in SEG literature or proposed in this paper. For instance, current developments have focused on improving actuator design to tackle more DOF [67]. During SEG assembly, the actuators are mounted into the dorsal side of the hand to avoid finger movement obstruction [68] and can be removed from the glove [69]. Actuators must not affect the active ROM of the finger joints and should allow free motions with more contact area for grasping tasks in a compliant manner [21].

Furthermore, actuators should take less than 4 s for full grasping [1]. The length of actuators should not be longer than the length of the fingers to avoid mismatching problems between them [23]. Actuators with low power consumption and continuous hours of operation are recommended.

2.2.2. Material Criterion

To enhance SEG operation, researchers continue to seek compliant, flexible, and lightweight materials to easily conform hand-finger anatomy with the shape of an object [41]. Hence, the payload capacity of elastomers has been exploited to obtain an elastic modulus similar to that of human tissues and avoid cumbersome designs [70].

Nonferromagnetic materials such as nylon, neoprene, polyester, or synthetic leather have been selected as compliant and affordable options to increase conformability and grip strength and reduce pressure on the skin [51, 71]. Additionally, silicon materials offer stable fastening and prevent slippage [72]. These synthetic polymers are easy to wash and do not absorb sweat compared to textile materials [23].

SEG made of fabrics have low cost and offer minimal mechanical impedance to finger motion [73]. Hence, coated fabric SEG systems with thermoplastic polyurethane (TPU) actuators are recommended for customization and to avoid slipping or muscle expansion problems [74].

Actuators made of fabrics work at lower pressures than elastomer actuators due to their inherent stiffness [75]. Therefore, several researchers have work on design, characterization, manufacture, and evaluation of soft elastomer actuators for hand [7678] and wrist [79] rehabilitation.

To match and support finger flexion-extension, some designs include multisegment elastomers with fiber reinforcement [80, 81] or corrugated fabric layers [41, 43] which are pressurized from 70 kPa up to 375 kPa [75]. Other designs include rigid plastic hoops [67] or nylon strings [82] to avoid radial deformations in FREA.

Material selection has also played a significant role in fastening the actuators to the glove or fingers in a safe way. Mostly, SEG proposals have employed magnets [83] or straps made of Velcro® [8, 18], fabrics [84], and rubber [24]. Other designs had opted for sewing the components [71] or separating the system from the actuators to reduce weight. Actuators can be attached to the wrist through elastomer bracelets [39] or synthetic hide covers [25, 31].

2.2.3. Manufacture Criterion

Mobility is also determined by manufacturing processes since specific elements can be obtained by particular methods that, additionally, can determine the weight and dexterity of the system. Conventional manufacturing procedures involve polymer casting molds [85], reinforcements and inclusions [11], additive manufacturing, thin-film manufacturing, shape deposition manufacturing, and bonding [86].

Mostly, 3D printing two-part mold has been employed for SEG spacers [23], cable guides [73], and elastomer actuators [87] where one mold is used to create a fluid chamber inside the actuators and the second one is addressed to create a fabric layer on top of the actuators [41, 43]. Nevertheless, low repeatability is the main drawback during this process [48].

Recent developments involve thermomethods [34], inverse flow injection [42, 82], lost wax molding [88], or fused deposition modeling with 3D printing at home to reduce SEG costs and facilitate its acquisition [89]. However, there is still room to improve SEG materials and fabrication with low costs.

New trends are oriented to hybrid designs where they combine rigid and soft components to obtain more hand poses and more DOF [90] and provide active training that encourages user participation [91].

2.2.4. Motion Guidance Criterion

SEG are designed to follow specific trajectories defined by a therapist depending on the impairment of the patient. These trajectories seek to achieve a functional ROM during both active and passive modes.

SEG is aimed at promoting active finger flexion and passive extension to increase patient autonomy during eating or drinking tasks [46, 92]. In the active assistance mode, patients attempt to move their hand and SEG are an additional aid to complete the desired ROM [93] whereas in the passive assistance mode, the exoglove provides all the assistance to guide the desired movement [94]. In the patients’ force recovery processes, effective SEG systems should, actively, participate with intensive training based on active and repetitive practical motions [95].

SEG should combine active and passive mobilizations for successful hand rehabilitation. For example, authors in [15] provide active extension on each finger. In [8], SEG also exert passive extension with active flexion and thumb opposition for grasping tasks. Other systems include active finger adduction-abduction [85] and perform flexion-extension motions [39, 71]. More sophisticated SEG systems have already begun an age that allows patients to perform a desired movement. When patients are able to achieve functional ROM, the system will have no effect on the hand [41] or will create an opposite force to improve the power of the patient.

Most of the reported SEG systems focus on PAM, a few on AAM as well as on the combination of active and passive modes (see Figure 2).

2.2.5. Operation and Control Criterion

SEG operation is defined by their type of actuation and their components. Tendon-driven wires require servomotors, gearboxes, spools, and force/torque sensors to move them. Pneumatic systems require a compressor, electrovalves or proportional valves, pressure sensors, or regulators. All these components are controlled on a data acquisition board which is plugged to a PC or uses Bluetooth as a communication interface for the SEG system [74].

Different schemes have been proposed to operate and control SEG systems; for example, in [14], Faulhaber 1226 006B motors, CompactRIO board, and LabVIEW® are used. Authors in [21] use DCX22 motors, a control board TMS320F2808®, and Simulink®. Additionally, graphical user interfaces (GUI)® have been implemented as a communication channel for SEG systems [88, 91]. A broad range of operation and control possibilities exists to select microcontrollers and interfaces relying on desired real-time response, accuracy, number of components involved in the operation and control strategies, and specific requirements of each SEG system.

Normally, open-loop control (OLC) and closed-loop control strategies are implemented during SEG operation. OLC schemes have used springs [1, 34] or mechanical switches [54] for manual operation where patients are able to drive an actuator to accomplish a specific task [96]. OLC strategies require the system to be stable by construction. To regulate the desired variables or to track specific trajectories that ensure patient safety while using a soft exoglove, CLC strategies are implemented [97]. To achieve acceptable motions in CLC schemes, sensors are directly attached to SEG [98, 99] without the patient worrying about making accurate movements.

Proportional (P) [68], proportional-derivative (PD) [100, 101], or proportional-integral-derivative (PID) [15, 71] controllers are widely implemented for flow and force regulation. Pulse width modulation (PWM) signals have been used to open and close solenoid valves [51] and can be implemented in many control strategies for different applications. Other kinds of controllers can be used depending on the system nature and on the task objective. For instance, nonlinear controllers, fuzzy approaches, or optimal linear control schemes could be developed for specific SEG systems. For instance, [102] provides an interesting review of soft robotic manipulator control strategies that could be considered to be applied in SEG systems.

SEG operation is based on force and position requirements to emulate human hand functions. These requirements, among others, are taken into account to define the control strategy to be synthesized. For example, SEG should have less than 10 minutes of setting time to become a useful tool for therapists [103]. Regarding fluid actuation, 10 N to 15 N are required for grasping tasks [11, 41]. SEG must be able to generate 7 N per finger or around 25 N on the whole hand with distributed forces along the fingers to minimize pressure location points, according to [34]. Normally, actuators with variable stiffness require 120 kPa for pinching and 160 kPa for grasping [18] while multisegment actuators require between 345 kPa and 400 kPa for flexion motion [51]. Desired joint ROM define positions to be reached by the patient when using a SEG system and provide reference variables to be controlled.

To evaluate SEG effectiveness in rehabilitation or assistance tasks, surface electromyography (EMG) has been implemented to detect user movement intentions [53], point out muscle contractions [16], control finger motion, and force level activation of muscles [90] since this is a noninvasive procedure that prevents muscle injuries.

During gripping tasks for finger flexion-extension, EMG signals are captured from the extensor digitorum communis (EDC) muscle together with the flexor digitorum superficialis (FDS) [20, 50] or with the flexor digitorum profundis (FDP) muscle [43, 73] since these muscles have been used and tested to work properly when implementing EMG procedures and due to the number of fingers they are connected with. Then, data obtained from a set of electrodes are amplified, filtered, quantified, and converted from analog to digital signals during SEG use [104]. This electrical stimulation should be monitored at least every 10 minutes to avoid muscle fatigue [103]. EMG signals can be used as control inputs when it is required to move specific hand joints that are connected to the aforementioned muscles. Due to stable behaviours, force myography (FMG) signals have been proposed to control the intention of the movement on SEG systems [20].

Motor impairment scales are applied to evaluate patient ROM to determine SEG operation ranges before starting an aided rehabilitation process. These scales serve for the evaluation of the damage that each patient has. According to [38, 105], patients with an Ashworth spasticity index (ASI) value less than or equal to three can use a soft exoglove. A modified Ashworth scale (MAS) value less than or equal to two defined the use of a soft exoglove for active flexion-extension, according to [106]. SEG operation is also based on the functional independence measure (FIM) of the patient by which the value goes from 1 to 7 depending on the assistance intensity [45]. Thus, for values above 3, patients present more autonomy [36].

2.2.6. Assessment Criterion

To ensure patient safety and SEG operation, several tools such as joint contractions [31, 54], bending angles [71], 3D visual motion analysis [11], or optical ROM at specific joints [44] have been employed to evaluate SEG performance. Other methods have opted for using mathematical models together with the finite element method (FEM) for hand and finger trajectory characterization [67]. To assess patient satisfaction when using SEG systems, questionnaires have been considered [88].

SEG assessment can be also done based on the blocked [48], grasping [21], pinching [44], or fingertip [14] forces that are quantified using bottles, cups, balls, telephones, cans, or fruits with variable mass, size, and texture [44, 51]. For cylindrical objects, the diameters go from 50 mm to 120 mm [21, 75] with a mass of 300 g [107]. Experimental tests on SEG assessment have been carried out with dummy hands [71] and healthy individuals [44] or combining healthy people and stroke survivors [75]. Other SEG evaluations perform tasks with/without a soft exoglove and compare them [46, 92]. ROM data have been collected when using a soft exoglove and without it [31].

To assess hand function and ROM using SEG systems, patients undergo coordination and dexterity tests. For example, the Kapandji score is used to evaluate thumb performance on pinching and grasping tasks [108]. SEG assessment also considers the motricity index test (MIT) [105], the Fugl-Meyer assessment (FMA) [46], the nine-hole peg test (NHPT) [38], the Jebsen-Taylor hand test (JTT) [44], the box and block test (BBT) [11], the Purdue pegboard test (PPT) [45], or some writing tasks [109].

For each patient, one or more of the aforementioned methods could be chosen by his/her motor impairment or by the therapist in charge of the respective rehabilitation protocol in order to assess SEG systems.

Some authors have focused more on statistical analysis about user condition than SEG performance [5, 40]. They seek for a specific target group, rehabilitation time, training tools, age, or gender, for instance.

2.3. SEG Usability Criteria

To guarantee a friendly and comfortable SEG use, modularity, portability, customization, mode of intervention, and cost criteria must be considered to develop a soft exoglove with particular characteristics as easy to put on and operate, working in an intuitive way, and having low cost. These criteria are discussed below.

2.3.1. Modularity Criterion

SEG designs have opted for modular configurations to ease donn and doff as in the cases of [21, 72, 83]. Connections can be assembled to work on targeted tasks, and actuators are mounted one by one [39]. Besides, modular designs for bending motions with deployable mechanisms have been adopted to reduce weight and allow natural motion [48]. SEG quality can be improved by a modularized system with relatively low cost customization, easy maintenance, and low power consumption [23]. Additionally, modular architectures allow for the replacement of feasible SEG components [89]. Based on this information, modularity is highly recommended as one of the main characteristics of SEG systems.

2.3.2. Portability Criterion

To cope with patients’ demands and to guarantee continuous rehabilitation protocols, the use of SEG outside clinics has become a main design concern to foster external rehabilitation [38, 110]. Nevertheless, to achieve this objective, SEG performance depends on the number of hours they can operate continuously without having a fixed power supply. According to [51], an effective soft exoglove should achieve, without problems, 2 hours of continuous operation or from 4 to 6 hours of intermittent operation.

Moreover, the runtime of batteries should be more than one hour in order to guarantee the development of a rehabilitation protocol session [100] until its completion or exert from 15 to 20 minutes of passive guidance [88]. Normally, lithium-polymer batteries are used since they can last 3.8 hours of continuous operation [23, 51].

Patients should take physical therapy sessions at rehabilitation facilities as well as at home [46, 92] in order to perform exercises on their own and not only depend on the availability of therapists [34]. SEG must be lightweight to allow their transportation [31, 73]. Thus, control unit boxes should be set up independently of the glove to minimize additional load [74]. Some power supply designs include waist belts [43, 51, 89], backpacks [73], boxes [11, 50], vests [84], waist pockets [53, 59], pockets [44], or a separate section located on another part of the human body [25, 31].

2.3.3. Customization Criterion

As established by [18], conformability, adaptability, and customization are some features that can be taken into account to fit, properly, the hand of a patient and generate a compliant soft exoglove. Particularly, customization affects SEG operation since each finger length varies due to sex, age, and finger palm size [17]. Thus, fasteners [71] and Velcro® straps [8, 110] have been used to attach, conveniently, SEG to hands. Otherwise, deviations from a nonappropriate size or form may restrict hand movement or cause discomfort during SEG use [21].

2.3.4. Mode of Intervention Criterion

To increase hand function rehabilitation, a bilateral mode in SEG systems results more beneficial than unilateral mode since patients can integrate healthy and paretic hand motions during rehabilitation therapy [75]. The bilateral mode is supported by a master-slave therapy concept where healthy limbs act as masters and soft devices as slaves [101]. Then, healthy limbs become a support for paretic limbs whereas devices working in the unilateral mode only exercise the impaired limb [111]. Bilateral mode rehabilitation could be recommended by the therapist as a function of the impairment. Then, SEG design could consider the mode of intervention depending on the associated rehabilitation protocol.

2.3.5. Cost Criterion

It has been noted that researches are more interested in the functionality of their products than in their price, since only few works report SEG costs. However, SEG cost will determine one of the aspects for the success of an exoglove as a commercial product. Therefore, designers could generate low-cost readily available SEG systems. For instance, authors in [34] propose that the assembly should cost less than $30 USD in order to be a competitive choice. Another proposal establishes that manufactuing and electronics should be less than $200 USD [100]. According to [52], soft exosuits for the upper limb should cost less than $1000 USD, $465 USD for the elbow, and $470 USD for the hand. A detailed description about the component cost of these configurations could be found in [59]. SEG costs can vary due to the type of actuation, the type of components and materials, the weight, and the country where they were developed [112].

Remarkable results about cost analysis between conventional and aided therapy show that SEG rehabilitation is more affordable than therapist assistance since the reported cost associated with aided therapy is almost three times less expensive than the conventional one [45].

Currently, Neofect™, Glohera™, and Bioservo™ companies have already patented their systems which have been commercially exploited for hand rehabilitation and assistance. However, these commercial systems are available only in some countries and are, relatively, expensive. Importation and shipping costs must be added to final prices for countries and locations where these systems are not available.

3. SEG Design Guidelines

Based on the information provided in Section 2, Table 2 summarizes some of the main aspects related to the 13 proposed design criteria for SEG developments.


TypeCriteriaConsiderations

FunctionRehabilitationStroke survivors and hand disabilities
AssistanceGrasp, grab, pinch, lift, hold, and release tasks

MobilityActuationCable-driven, pneumatic, or hydraulic
MaterialsFabric, synthetic leather, neoprene
Motion guidancePassive or active
Manufacture3D printing+other procedures
Operation & control45 minutes/day, OLC or CLC
Assessment7 N/finger or 10 to 15 N for 1 kg objects

UsabilityModularityOpen or closed palm configuration
Portability500 g or less than 3 kg for the whole system
CustomizationRight size, fasteners or Velcro® straps
Mode of interventionBilateral or unilateral
CostsAssembly less than $30 USD

At present, SEG approaches are focused on improving functionality, strength, DOF, and ROM for object manipulation. Figure 2 and Table 3 provide information for each of the 91 SEG systems reviewed in this paper, associated with the following 15 aspects: (1) function: robot rehabilitation (RT) or assistance tasks (AT); (2) application: hand disability, stroke survivors, or SCI; (3) task: grasping, pinching, or manipulation; (4) palm design: OPD or CPD; (5) type and number of actuators: tendon-driven, pneumatic, or hydraulic; (6) assistance mode: PAM or AAM; (7) DOF per finger; (8) targeted fingers; (9) motions: flexion-extension, adduction-abduction, opponent, ulnar/radial deviations, and pronation-supination; (10) material; (11) weight; (12) force; (13) pressure; (14) control: CLC or OLC; and (15) assessment.


SEG # (Ref.)FunctionApplicationTaskPalm designActuation (number)Assistance modeDOF/fingerFingersMotionMaterialWeight (g)Force (N)Pressure (kPa)ControlAssessment

1 [14]ATStroke survivors, SCIGraspingCPDTendon-driven (1)PAM8Index, middle, thumbFlexion-extensionSynthetic latex8018CLCFingertip and blocked force
2 [17]AT-RTHand disabilityGraspingCPDTendon-driven (3)PAM8Index, middle, thumbFlexion-extensionSynthetic latexCLC (PI, PD)EMG
3 [19]ATStroke survivors, SCIGraspingCPDTendon-driven (3)PAM8Index, middle, thumbFlexion-extensionSynthetic latex19420CLCEMG
4 [21]ATHand disabilityGraspingOPDTendon-driven (2)PAM8Index, middle, thumbFlexion-extensionSilicone KE-130020EMG
5 [23]ATHand disabilityGrasping, pinchingOPDPneumatic (4)PAM11Index, middle, thumbFlexion-extension, adduction-abductionSilicone KE-1300 T35022.5300CLCROM, grip strength
6 [22]ATSCIGrasping, pinchingOPDTendon-driven (3)PAM9Index, middle, thumbFlexion-extension, opponentSilicone KE-1300 T10410.3CLCGrip strength
7 [8]RTHand disabilityGrasping, pinchingOPDPneumatic elastomer (4)PAM14All except thumbFlexion-extension, curlNeoprene160CLCFEM analysis
8 [51]AT-RTHand disabilityGrasping, pinchingOPDHydraulic elastomer (2)PAM15AllFlexion-extension, opponentNeoprene400CLCROM, EMG
9 [11]AT-RTSCIGraspingOPDHydraulic elastomer (2)PAM15AllFlexion-extension, opponentTextile10-15CLCROM, BBT
10 [50]ATHand disabilityGrasping, holdingOPDHydraulic elastomersPAM15AllFlexion-extension, opponentElastomer14413CLCROM, EMG
11 [36]RTStroke survivorsGrippingOPDPneumaticPAM16AllFlexion-extension, opponent, adduction-abductionROM: MI, BBT, FIM
12 [105]AT-RTHemiplegic patientsGripping, pinchingOPDPneumaticPAM16AllFlexion-extension, opponent, adduction-abductionROM: MI, NHPT test
13 [45]AT-RTStroke, SCIGripping, pinchingOPDPneumaticAAM-PAM16AllFlexion-extension, opponent, adduction-abductionROM: NHPT, FIM test
14 [113]RTStroke, SCIGripping, pinchingOPDPneumaticAAM-PAM15AllFlexion-extension, opponent, adduction-abductionROM: MI, NHPT test
15 [38]RTStroke survivorsGrippingOPDElectrical (5)PAM15All, wristFlexion-extension, opponent, adduction-abductionAshworth index
16 [114]ATGraspingCPDHydraulicPAM14AllFlexion-extension262012550CLCPressure regulating
17 [39]RTStroke survivors, SCIGrasping, manipulationOPDTendon-drivenAAM-PAM14All, wrist, forearmFlexion-extension, opponent, pronation-supinationROM: FMA, JTT, PPT
18 [115]RTStroke survivorsGrasping, manipulationOPDTendon-drivenAAM14All, wrist, forearmFlexion-extension, opponent, pronation-supination
19 [47]RTStroke survivors, SCIGrasping, manipulationOPDTendon-drivenAAM-PAM14All, wrist, forearmFlexion-extension, radial-ulnardeviations, pronation-supinationElastomer132CLCROM
20 [73]AT-RTStroke survivors, SCIGrasping, pinchingCPDTendon-driven (5)AAM-PAM12AllFlexion-extensionLycra, fabrics15CLCGrip force
21 [84]ATHand disabilityGrasping, pinchingCPDTendon-driven (3)PAM8Middle, ring, thumbFlexion-extensionSynthetic leather70020CLC (PID)Grasping power test
22 [85]AT-RTHand disabilityGraspingOPDTendon-driven (5)PAM15AllFlexion-extension, adduction-abductionSilicone rubber17.25165CLCROM: tactile pressure
23 [116]RTHand disabilityGraspingOPDTendon-driven (5)PAM15AllFlexion-extension, adduction-abductionElastomer
24 [24]ATMaterial handlingGraspingCPDTendon-drivenPAMAllFabricsCLCLift forces
25 [18]AT-RTHand disabilityGrasping, pinchingCPDPneumatic (5)AAM14AllFlexion-extensionFabrics200160CLCROM
26 [41]RTStroke survivorsGraspingCPDPneumatic (4)PAM12All except thumbFlexion-extensionFabrics2009.25200CLCfMRI, optical ROM
27 [43]ATGrasp pathologiesGrasping, releasingCPDPneumatic (5)PAM15AllFlexion-extension, opponentLycra, fabrics17013.6153CLCOptical ROM, EMG
28 [1]AT-RTStroke survivorsGrasping, releasingOPDPneumatic (5)AAM14AllFlexion-extensionNeoprene150100CLCROM, torque
29 [106]AT-RTStroke survivorsGrasping, pinchingOPDPneumatic (4)PAM12All except thumbFlexion-extensionLycra, fabrics18010.2120CLCfMRI, optical ROM
30 [44]ATStroke survivorsGrasping, lifting, releasingOPDPneumatic (5)AAM-PAM14AllFlexion-extensionLycra, fabrics18012-36120CLCROM, EMG
31 [75]AT-RTHand disabilityGrasping, manipulationCPDPneumatic (5)AAM14AllFlexion-extensionFabrics9913.6275-375CLCGripping force
32 [53]ATHand disabilityGraspingCPDTendon-driven (1)PAM8Index, middle, thumbFlexion-extensionSynthetic latex50010CLCROM, gripping force
33 [59]ATMuscle weaknessGraspingCPDTendon-driven (2)PAM8Index, middle, thumb, elbowFlexion-extensionNeoprene120010CLC (PD)ROM, gripping force
34 [52]ATHand disabilityGraspingCPDTendon-driven (3)PAM8Index, middle, thumbFlexion-extensionNeoprene500CLC (PD)ROM, gripping force
35 [25]AT-RTHeavy tasksGrasping holdingCPDTendon-driven (3)PAM14AllFlexion-extensionFabrics770CLCGrasping force
36 [30]AT-RTHand disabilityGrasping, manipulationOPDPneumatic, hybrid (4)PAM12All except thumbFlexion-extensionNylon1502.5230CLCBending forces, EMG
37 [48]RTHand disabilityGrasping, manipulationCPDPneumatic, hybrid (5)PAM14AllFlexion-extension165.4CLCROM, joint angles
38 [42]RTHand pathologiesGraspingOPDPneumatic elastomers (5)PAM14AllFlexion-extensionTextile526CLCROM, fatigue test
39 [16]ATHand disabilityGrasping, pinchingCPDPneumatic (5)PAM15AllFlexion-extension, opponentSynthetic leather1359200CLCROM, EMG
40 [15]RTStroke survivorsGrasping, releasing, pinchingCPDPneumatic (5)PAM14AllFlexion-extensionLycra1560CLC (PID)ROM: FMA, BBT
41 [31]RTHand disabilityGraspingCPDTendon-driven (3)PAM8Index, middle, thumbFlexion-extensionSynthetic leatherCLCROM, joint angles
42 [35]AT-RTHand paralysisGraspingCPDTendon-driven (1)PAM8Index, middle, thumbFlexion-extensionPolyester fiber5035CLCROM, EMG
43 [117]AT-RTHand disabilityGrippingCPDTendon-driven (3)PAM8Index, middle, thumbFlexion-extensionFabricsCLCROM, gripping force
44 [118]ATOlder adultsGrippingCPDTendon-driven (3)PAM8Index, middle, thumbFlexion-extensionFabrics85CLCPinch strength, JTHFT
45 [46]AT-RTHand disabilityGrippingCPDTendon-driven (5)PAM14AllFlexion-extensionCLCPinching force, JTHFT test
46 [54]RTHand disabilityGraspingCPDTendon-driven (5)PAM14AllFlexion-extensionFabricsCLCROM, joint angles
47 [71]RTStroke survivorsGraspingOPDPneumatic (5)PAM16AllFlexion-extension, opponent, adduction-abductionNylon20CLC (PID)ROM, joint angles
48 [34]RTStroke survivorsGraspingCPDSpring mechanism (5)PAM14AllFlexion-extensionSynthetic leather20022.59Bending force
49 [110]RTStroke survivorsGraspingOPDPneumatic (5)PAM14AllFlexion-extensionCLCElectroencephalography
50 [89]AT-RTStroke survivorsGrippingOPDPneumatic (5)PAM14AllFlexion-extensionElastomer41.8200CLCEMG, ROM, gripping force
51 [107]ATHand disabilityGrasping, manipulationCPDCable-driven (4)PAM11All except littleFlexion-extension25016CLCROM, pinching force
52 [67]RTHand disabilityGraspingCPDPneumatic (4)PAM12All except thumbFlexion-extensionNylon3200FEM & ROM
53 [119]ATHand disabilityGrasping, manipulationOPDPneumatic artificial muscles (5)PAM14AllFlexion-extensionFabrics16110200FEM, fingertip force
54 [120]AT-RTOlder adultsGrasping, manipulationOPDTendon-driven (5)PAM14AllFlexion-extensionTPU, NINJAFLEX™504060CLCPressure regulation & fingertip force
55 [74]AT-RTHand disabilityGrasping, pinchingOPDTendon-driven (5)PAM14AllFlexion-extensionTPU33022 pinch, 48 graspCLCPinching and grasping forces
56 [55]AT-RTHand disabilityGraspingCPDShape memory alloys (5)PAM14AllFlexion-extensionFabrics40CLCROM, fingertip-tendon force
57 [4]ATStroke survivorsGrasping, manipulationOCPTendon-driven (5)PAM12All except thumbFlexion-extensionPolymer340CLCGripping force
58 [112]ATCP, stroke survivorsGraspingCPDTendon-driven (3)PAM8Thumb, index, and middleFlexion-extensionSynthetic leather5548CLCGrasping and fingertip forces
59 [121]ATSupportive aidGraspingCPDShape memory alloys (5)PAM12All except littleFlexion-extension, radial abduction, palmar abduction, oppositionSynthetic leather85.0311CLCGrasping force, ROM
60 [122]RTStroke survivorsGraspingOPDPneumatic (1)PAM3IndexFlexion-extensionEcoflex™ 00-301.1730CLCBending angle
61 [13]ATOlder adults, hand disabilityGrasping, pinchingCPDPneumatic artificial muscles (5)PAM14AllFlexion-extensionRubber5.7-14, 20-25500CLCGrasping and pinching forces
62 [109]ATSCIGraspingCPDTendon-driven (3)PAM8Thumb, index, and middleFlexion-extension7.39CLCWriting tasks, grasping force
63 [7]AT-RTStroke survivorsGraspingCPDPneumatic, tendon-driven (5)PAM14AllFlexion-extensionNylon5300CLCAshworth test
64 [123]RTStroke survivorsGraspingCPDTendon-driven (4)PAM12All except thumbFlexion-extensionCLCFMA test
65 [82]RTHand disabilityGraspingOPDPneumatic FREAs (5)PAM14AllFlexion-extensionNylon28011.27250CLCGrasping test FEM
66 [108]RTHand disabilityGraspingOPDPneumatic FREAs (1)PAM1ThumbOppositionElastomer586150CLCKapandji test
67 [124]ATStroke survivorsGraspingOPDPneumatic FREAs (5)PAM14AllFlexion-extensionSilicone rubber207200CLCBending force FEM
68 [125]ATSCIGrasping, manipulationOPDPneumatic (5)PAM14AllFlexion-extensionFabric7715172CLCLifting force
69 [99]ATStroke survivorsGraspingCPDElectricalPAM15All, wristFlexion-extensionNeoprene fabricCLCBBT test
70 [98]RTStroke survivorsGraspingCPDTendon-driven (1)PAM3Index, wristFlexion-extensionLycraCLCROM
71 [61]RTKinesthetics, haptic feedbacksPressingOPDPneumatic (2)PAM6Index and middleFlexionSilicone rubber16.66210CLCVirtual reality haptic feedback
72 [126]ATHand disabilitiesGrasping, holdingCPDPneumatic (4)PAM15AllFlexion-extension, opposition65CLCGrabbing force
73 [57]ATHand disabilitiesGraspingCPDPneumatic (4)PAM11All except littleFlexion-extension16025500CLCBending angle
74 [127]ATHand disabilitiesGraspingCPDPneumatic (5)PAM14AllFlexion-extensionElastomer1803300CLCGrasping force
75 [68]RTStroke survivorsGraspingCPDPneumatic (5)PAM14AllFlexion-extensionLatexCLC proportionalBending angle
76 [72]RTStroke survivorsGraspingOPDTendon-driven (5)PAM9Thumb, index, and middleFlexion-extension, opposition/repositionSilicone KE-1300 T12012CLCBending angle
77 [49]ATStroke survivorsGraspingCPDTendon-driven (5)PAM12All except thumbFlexion-extensionLycra16-17CLCFingertip force ROM
78 [100]RTStroke survivorsGraspingCPDTendon-driven (5)AAM-PAM14AllFlexion-extensionElastomer>1000CLC (PD)Fingertip force ROM
79 [69]AT-RTHand disabilitiesGraspingCPDPneumatic FREA (5)PAM16AllFlexion-extension, opponent, adduction-abductionPolyester760.8150CLCBending angle and force output
80 [88]RTStroke survivorsGraspingOPDPneumatic (5)PAM14AllFlexion-extensionRTV-4234T4, silicon50CLC (PD)Bending angle
81 [40]RTStroke survivorsGraspingCPDTendon-driven (5)PAM14AllFlexion-extensionFabricsVirtual reality, FMA
82 [10]ATHand disabilitiesGraspingSemiopenMotor-tendon (5)PAM14AllFlexion-extensionCotton fabric600On-off controlGrasping force output
83 [90]ATHand disabilitiesGraspingCPDTendon-driven (5)PAM12All except littleFlexion-extension, opponent22083CLCGrasping force ROM
84 [26]ATHeavy tasksManipulation tasksCPDTendon-driven SMA (5)PAM14AllFlexion-extensionRubber70CLC (PID)Force output
85 [104]ATHand disabilitiesGrasping, releasingCPDPneumatic (5)PAM14AllFlexion-extensionFabric16088.29180EMG signals
86 [91]RTHand disabilitiesGraspingCPDSteel springAAM-PAM14AllFlexion-extension40130.87CLCROM, force output, EEG signals
87 [58]RTHand disabilitiesGraspingCPDPneumatic (5)PAM14AllFlexion-extension15040300OLCEMG signals, grasping forces
88 [96]ATStroke survivorsGraspingCPDTendon-driven (5)AAM-PAM10AllFlexion-extensionNylon258OLCEMG signals, grasping and lifting forces
89 [103]RTStroke survivorsGraspingCPDTendon-driven SMA (5)PAM14AllFlexion-extensionTeleoperation, time output
90 [101]RTStroke survivorsGraspingCPDPneumatic (5)PAM14AllFlexion-extensionRTV-4234T4, silicon105CLC (PD)Finger trajectories and angle
91 [20]ATHand disabilitiesGraspingCPDTendon-driven (5)PAM14AllFlexion-extensionPolyester and neopreneCLCGriping force and FMG signals

Figure 2 provides information related to the number of soft exogloves that have been developed in the last decade, being characterized by particular aspects. For example, the most important number of SEG systems that have been developed is focused on the passive assistance mode, CLC predominate over open-loop strategies, elastomers are preferred to other types of material, hydraulic actuation is not significant compared to the number of SEG devices using tendon-driven or pneumatic actuation, and SEG have been developed, mainly, to cope with stroke and hand disabilities as well as with rehabilitation and assistance problems.

Based on what has been presented so far, the following SEG design guidelines are highlighted in order to be considered when developing new SEG systems. (1)Rehabilitation and assistance tasks should be included in a single soft exoglove(2)SEG are primarily designed for stroke survivors and people with hand disabilities(3)Grasping is the main assistance task that has been addressed by SEG systems(4)SEG have been diversified for both OPD and CPD depending on the actuation(5)Tendon-driven and pneumatic are preferable types of actuators(6)AAM should be the priority motion guidance for SEG rehabilitation(7)Mostly SEG provide more than 10 DOF to reach hand motor function(8)A complete hand characterization must be included to tackle more DOF(9)All SEG provide, at least, flexion-extension motion. Furthermore, adduction-abduction and opponent motions are desirable(10)Elastomers have become the main material choice due to their flexibility, lightness, and adaptability(11)SEG systems should have a total mass of less than 200 g to enhance their efficiency(12)SEG should provide, at least, 5 N per finger to execute most of ADL(13)Regarding pneumatic actuation, SEG should work between 100 and 300 kPa(14)CLC controllers are preferable to OLC in order to ensure patient safety and system precision. Particularly, PD controllers have been mostly implemented(15)Fingertip forces, ROM, and EMG are the most used tools to evaluate SEG effectiveness

Table 3 provides detailed information related to the 15 aspects illustrated in Figure 2 for 91 devices that have been analysed in order to identify, classify, and discuss the 13 aforementioned criteria and to establish the previous 15 design guidelines for SEG systems. For example, the third system has eight DOF, focuses on grasping assistance tasks, has a closed palm configuration, is passively driven (CLC) by cables, and performs flexion/extension of 3 fingers.

From the previous information reported in this paper, five core SEG developers have been identified and have marked trends in the design of soft exoglove systems. Hong Kai Yap is the author with the highest number of SEG contributions (see Table 3, items 25-31).

The number of SEG developments, from the last ten years, is plotted in Figure 3. According to literature, 2017 was the most productive year with 21 of the 91 contributions reported in this paper.

4. Discussion

In order to provide technical solutions for hand rehabilitation and assistance, multiple endeavours have been done during the last three years about SEG developments [128].

This review has identified areas of opportunity for the improvement of soft exogloves that are used in aided rehabilitation protocols and assistance tasks. Four main circumstances have motivated researchers to satisfy popular demand and increase SEG development since they represent an alternative and affordable approach to overcome hand disabilities. These circumstances are related to the increase in the number of people with hand motor deficits, to poor therapist availability, to the fact that clinical facilities are struggling to provide rehabilitation training, and to the expensive costs of these health services.

There are still significant challenges to face in soft exoglove design. For instance, power supply approaches are still limited and tendon-driven actuation necessitates motors without heating problems, whereas hysteresis issues should be solved in pneumatic systems to increase actuation cycles and durability along with lightweight and portable power supplies.

Regarding rehabilitation approaches, SEG systems must be endowed to exert intensive and repetitive routines without muscle fatigue and with minimal therapist assistance to excel above other rehabilitation options. SEG are a supportive aid that contributes to accelerated hand recovery by therapy protocols. Nevertheless, to achieve a desired rehabilitation task, an active contribution from the patient is required to regain strength, mobility, and ROM. Since the progress of each patient is variable, an AAM with time-triggered control could be implemented to regulate the input force of patients during rehabilitation processes, depending on their physical condition. SEG systems must encourage patient participation but do not execute all the rehabilitation work.

Several works have demonstrated that soft exogloves have the potential to offer safe human-robot rehabilitation or assistance. However, new trends show that these two tasks should be integrated into a unified system as it is reported by [46, 92]. To accomplish integral rehabilitation, SEG designers must consider that modular devices are expected to help therapists and patients depending on the impairment or on the rehabilitation protocol. This will be satisfied by connecting a soft exoglove device to a soft exosuit with a reliable and robust platform (see, for instance, [28]).

SEG shortcomings were identified concerning different hand sizes since most available systems are oriented towards adults. Thus, adjustable devices are recommended to have the possibility to initiate an early SEG-based rehabilitation program since this is a common advice given by therapists, no matter the dimensions of the patient’s hand. So far, SEG systems are able to accomplish full open-close fist, grasping, lifting, and object release. Therefore, the systems reported in literature encompass from 8 to 14 DOF. Moreover, SEG characterization could be developed to obtain more DOF in order to expand the workspace if needed.

When soft exogloves are used, patient safety is a priority. Thus, human-machine interfaces with emergency buttons and haptic feedback must be considered for harmless interactions [35, 128] as stated in Section 2 of this paper, and several safety strategies must be incorporated in every SEG system. Moreover, SEG systems should not obstruct natural hand mobility and do not affect active ROM. Additionally, new developments are expected to provide patients and therapists with useful information in order to evaluate patient progress. Furthermore, the capability to automatically adjust the operation parameters as a function of the patient recovery level is desirable.

SEG self-manufacturing designs must ensure functional operation for home rehabilitation to provide low-cost systems. These considerations could allow to improve SEG features as hours of operation, power consumption, cleaning, and maintenance. Since Bluetooth communications have been considered between SEG systems and control interfaces [74], this or other communication systems must be part of new SEG devices when dealing with CLC strategies and for rehabilitation or assistance data analysis.

From this review, it can be pointed out that in recent years, the development of SEG has grown significantly in rehabilitation clinics and research groups. However, there is no comparison between research prototypes and those that have been already commercialized because the level of their technological maturity is different for each of them. Commercialized SEG systems must have evolved from research prototypes. The main difference between these two types of devices is the one related to their technological maturity. For instance, research prototypes can reach, in favorable cases, a technology readiness level (TRL) of 4 or 5 while commercialized products have the highest TRL of 9 in China [129, 130]. The evolution of a research prototype going from a 5 TRL to a certified product with 8 TRL and to a commercial product with a 9 TRL can take several years and require significant quantities of money. Moreover, medical devices having official approvals or certifications as that of the Food and Drug Administration (FDA) or the Conformité Européenne (CE) can be commercialized since they satisfy specific requirements and standards while research prototypes focus, mainly, on satisfying functional aspects. Then, it can be stated that commercialized medical devices are reliable due to the fact that they have completed the product design cycle reaching the product life-cycle management, while research prototypes have not begun the product development cycle or their industrial manufacture yet.

New-generation products should seek for an affordable trade-off between cost and benefit and include the possibility to perform assistance or rehabilitation therapy at home or in specialized clinics to ensure that rehabilitation protocols, defined by therapists, are efficiently executed.

SEG designs should provide acceptable appearance, comfort, and functionality to patients. Hence, it is highly recommended that SEG systems consider accessible technologies that could, additionally, create dynamic environments where patients can have pleasant therapy sessions. SEG require materials with appearance and elastic modulus similar to human tissues. Thus, smart polymers represent the primary current choice due to their biomimetic qualities to develop lightweight devices with modular OPD [128]. Besides, elastomers have been shown to be compliant wearable components with the ability to vary their form and increase the ROM based on the shape of the human hand.

Modularity plays a significant role when dealing with maintenance aspects of SEG systems as well as with costs and should be considered in new SEG developments. Besides, modularity can play a significant role when dealing with rehabilitation of different fingers or DOF. Regarding portability in new SEG developments, minimizing the dependence of energy sources becomes a challenge that must be addressed by researchers and engineers.

It has become clear that a SEG device that allows adaptation (customization) to a larger number of patients without the need for component replacements will be preferable to another system that only works for a certain size of hands.

5. Conclusions

Scientific and technical communications concerning wearable SEG for hand rehabilitation and assistance tasks applied to stroke survivors or people with hand disabilities have been extensively reviewed and reported in this paper. SEG design criteria have been identified, classified, and established into 2 function, 6 operation, and 5 usability criteria.

This paper also provides 15 guidelines for SEG design, a detailed description of 91 SEG that have been analysed based on the aforementioned criteria, and a discussion that considers different aspects in order to enhance future SEG developments.

From this review, it is highlighted that patient safety should be a priority characteristic during SEG operation, and then, it should be guaranteed in every new SEG development. This goal can be achieved by working closely with a therapist, as recommended in [28], as well as incorporating safety in mechanical and electronical parts and in the programming of the SEG device. Moreover, safety standards have been referenced to be considered in every SEG development.

It has been remarked that several efforts have been made in terms of SEG designs. However, there is still room to improve these devices. Then, this paper provides suggestions on patient safety, functional and continuous operation, friendly interaction, feedback information, and materials.

Other areas to be explored include hybrid SEG systems where new assembly techniques ensure force transmission or the use of electroencephalography signals to monitor brain activity when SEG rehabilitation is performed. SEG systems should be able to combine passive and active assistance modes along with bilateral training to enhance recovery processes and to encourage patients. The mentioned SEG design criteria provide perfectible guidelines to improve their performance and represent a basis to develop SEG robust designs.

Abbreviations

AAM:Active assistance mode
ADL:Activities of daily living
ARM:Active resistive mode
ASI:Ashworth spasticity index
AT:Assistance tasks
BBT:Box and block test
CLC:Closed-loop control
CE:Conformité Européenne
CMC:Carpometacarpal
CP:Cerebral palsy
CPD:Closed palm design
DOF:Degrees of freedom
EMG:Electromyography
FDA:Food and Drug Administration
FEM:Finite element method
FIM:Functional independence measure
FMA:Fugl-Meyer assessment
FMG:Force myography
FREA:Fiber reinforced elastomer actuators
GUI:Graphical user interface
IP:Interphalangeal
JTT:Jebsen-Taylor hand test
MAS:Modified Ashworth scale
MCP:Metacarpophalangeal
MIT:Motricity index test
NHPT:Nine-hole peg test
OLC:Open-loop control
OPD:Open palm design
PAM:Passive assistance mode
PD:Proportional-derivative
PID:Proportional-integral-derivative
PT:Purdue pegboard test
ROM:Range of motion
RT:Robot rehabilitation
SCI:Spinal cord injury
SEG:Soft exogloves
SMA:Shape memory alloys
RTV:Room-temperature-vulcanizing
TPU:Thermoplastic polyurethane
TRL:Technology readiness level.

Conflicts of Interest

The authors declare that there is no conflict of interest regarding the publication of this paper.

Acknowledgments

The authors thank CONACYT and Universidad Autónoma del Estado de México for all their support. Funding supporting this research was provided by CONACYT (406476) and Universidad Autónoma del Estado de México (5015-2020CIA).

References

  1. H. K. Yap, J. H. Lim, J. C. H. Goh, and C.-H. Yeow, “Design of a Soft Robotic Glove for Hand Rehabilitation of Stroke Patients With Clenched Fist Deformity Using Inflatable Plastic Actuators,” Journal of Medical Devices, vol. 10, no. 4, 2016. View at: Publisher Site | Google Scholar
  2. V. L. Feigin, B. Norrving, and G. A. Mensah, “Global burden of stroke,” Circulation Research, vol. 120, no. 3, pp. 439–448, 2017. View at: Publisher Site | Google Scholar
  3. A. G. Thrift, T. Thayabaranathan, G. Howard et al., “Global stroke statistics,” International Journal of Stroke, vol. 12, no. 1, pp. 13–32, 2017. View at: Publisher Site | Google Scholar
  4. B. Cherian, C. Dominic, G. Vysakh, and K. R. Vishakh, “Exo-Glove: A Soft Wearable Robotic Hand for Stroke Survivors,” International Research Journal of Engineering and Technology, vol. 5, no. 5, pp. 3358–3364, 2018. View at: Google Scholar
  5. B. Radder, G. B. Prange-Lasonder, A. I. R. Kottink et al., “Home rehabilitation supported by a wearable soft-robotic device for improving hand function in older adults: A pilot randomized controlled trial,” PLoS One, vol. 14, no. 8, article e0220544, 2019. View at: Publisher Site | Google Scholar
  6. B. Radder, “General Introduction,” The Wearable Hand Robot, pp. 10–21, 2018. View at: Google Scholar
  7. A. Stilli, A. Cremoni, M. Bianchi et al., “AirExGlove — A novel pneumatic exoskeleton glove for adaptive hand rehabilitation in post-stroke patients,” in IEEE International Conference on Soft Robotics (RoboSoft), vol. 2018, pp. 579–584, Livorno, Italy, 2018. View at: Publisher Site | Google Scholar
  8. P. Polygerinos, S. Lyne, Z. Wang et al., “Towards a soft pneumatic glove for hand rehabilitation,” in 2013 IEEE/RSJ International Conference on Intelligent Robots and Systems, pp. 1512–1517, Tokyo, Japan, 2013. View at: Publisher Site | Google Scholar
  9. Z. Yue, X. Zhang, and J. Wang, “Hand Rehabilitation Robotics on Poststroke Motor Recovery,” Behavioural Neurology, vol. 2017, Article ID 3908135, 20 pages, 2017. View at: Publisher Site | Google Scholar
  10. R. Ismail, M. Ariyanto, T. Hidayat, and J. D. Setiawan, “Design of fabric-based soft robotic glove for hand function assistance,” in 2019 6th International Conference on Information Technology, Computer and Electrical Engineering (ICITACEE), pp. 1–5, Semarang, Indonesia, Indonesia, 2019. View at: Publisher Site | Google Scholar
  11. P. Polygerinos, K. C. Galloway, E. Savage, M. Herman, K. O’Donnell, and C. J. Walsh, “Soft robotic glove for hand rehabilitation and task specific training,” in 2015 IEEE International Conference on Robotics and Automation (ICRA), pp. 2913–2919, Seattle, WA, USA, 2015. View at: Publisher Site | Google Scholar
  12. D. Rus and M. T. Tolley, “Design, fabrication and control of soft robots,” Nature, vol. 521, no. 7553, pp. 467–475, 2015. View at: Publisher Site | Google Scholar
  13. T. Noritsugu, M. Takaiwa, and D. Sasaki, “Power assist wear driven with pneumatic rubber artificial muscles,” in 2008 15th International Conference on Mechatronics and Machine Vision in Practice, pp. 539–544, Auckland, New Zealand, 2008. View at: Publisher Site | Google Scholar
  14. H. In, K. J. Cho, K. Kim, and B. Lee, “Jointless structure and under-actuation mechanism for compact hand exoskeleton,” in 2011 IEEE International Conference on Rehabilitation Robotics, pp. 1–6, Zurich, Switzerland, 2011. View at: Google Scholar
  15. L. Connelly, Y. Jia, M. L. Toro, M. E. Stoykov, R. V. Kenyon, and D. G. Kamper, “A Pneumatic Glove and Immersive Virtual Reality Environment for Hand Rehabilitative Training After Stroke,” IEEE Transactions on Neural Systems and Rehabilitation Engineering, vol. 18, no. 5, pp. 551–559, 2010. View at: Publisher Site | Google Scholar
  16. Y. Kadowaki, T. Noritsugu, M. Takaiwa, D. Sasaki, and M. Kato, “Development of Soft Power-Assist Glove and Control Based on Human Intent,” Journal of Robotics and Mechatronics, vol. 23, no. 2, pp. 281–291, 2011. View at: Publisher Site | Google Scholar
  17. U. Jeong, H.-K. In, and K.-J. Cho, “Implementation of various control algorithms for hand rehabilitation exercise using wearable robotic hand,” Intelligent Service Robotics, vol. 6, no. 4, pp. 181–189, 2013. View at: Publisher Site | Google Scholar
  18. H. K. Yap, J. H. Lim, F. Nasrallah, J. C. H. Goh, and R. C. H. Yeow, “A soft exoskeleton for hand assistive and rehabilitation application using pneumatic actuators with variable stiffness,” in 2015 IEEE International Conference on Robotics and Automation (ICRA), pp. 4967–4972, Seattle, WA, USA, 2015. View at: Publisher Site | Google Scholar
  19. H. In, B. B. Kang, M. Sin, and K. J. Cho, “Exo-Glove: A Wearable Robot for the Hand with a Soft Tendon Routing System,” IEEE Robotics and Automation Magazine, vol. 22, no. 1, pp. 97–105, 2015. View at: Publisher Site | Google Scholar
  20. J. Fajardo, A. R. Neto, W. Silva, M. Gomes, E. Fujiwara, and E. Rohmer, “A wearable robotic glove based on optical FMG driven controller,” in 2019 IEEE 4th International Conference on Advanced Robotics and Mechatronics (ICARM), pp. 81–86, Toyonaka, Japan, Japan, 2019. View at: Publisher Site | Google Scholar
  21. B. B. Kang, H. Lee, H. In, U. Jeong, J. Chung, and K. J. Cho, “Development of a polymer-based tendon-driven wearable robotic hand,” in 2016 IEEE International Conference on Robotics and Automation (ICRA), pp. 3750–3755, Stockholm, Sweden, 2016. View at: Publisher Site | Google Scholar
  22. B. B. Kang, H. Choi, H. Lee, and K. J. Cho, “Exo-Glove Poly II: A Polymer-Based Soft Wearable Robot for the Hand with a Tendon-Driven Actuation System,” Soft Robotics, vol. 6, no. 2, pp. 214–227, 2019. View at: Publisher Site | Google Scholar
  23. S. S. Yun, B. B. Kang, and K. J. Cho, “Exo-Glove PM: An Easily Customizable Modularized Pneumatic Assistive Glove,” IEEE Robotics and Automation Letters, vol. 2, no. 3, pp. 1725–1732, 2017. View at: Publisher Site | Google Scholar
  24. H. Kazerooni, D. Fairbanks, A. Chen, and G. Shin, “The magic glove,” in 2004 in IEEE International Conference on Robotics and Automation, 2004. Proceedings. ICRA '04. 2004, vol. 1, pp. 757–763, New Orleans, LA, USA, USA, 2004. View at: Publisher Site | Google Scholar
  25. M. Diftler, C. A. Ihrke, L. B. Bridgwater et al., “RoboGlove - A Robonaut Derived Multipurpose Assistive Device,” in International Conference on Robotics and Automation, Hong Kong, China, 2014. View at: Google Scholar
  26. Á. Villoslada, C. Rivera, N. Escudero, F. Martín, D. Blanco, and L. Moreno, “Hand Exo-Muscular System for Assisting Astronauts During Extravehicular Activities,” Soft Robotics, vol. 6, no. 1, pp. 21–37, 2019. View at: Publisher Site | Google Scholar
  27. P. Ohta, L. Valle, J. King et al., “Design of a Lightweight Soft Robotic Arm Using Pneumatic Artificial Muscles and Inflatable Sleeves,” Soft Robotics, vol. 5, no. 2, pp. 204–215, 2018. View at: Publisher Site | Google Scholar
  28. J.-M. Dávila-Vilchis, J. C. Á. V. LAZ-Avilés, and A. H. Vilchis-González, “Design Methodology for Soft Wearable Devices—The MOSAR Case,” Applied Sciences, vol. 9, no. 22, p. 4727, 2019. View at: Publisher Site | Google Scholar
  29. M. Troncossi, M. Mozaffari-Foumashi, and V. Parenti-Castelli, “An original classification of rehabilitation hand exoskeletons,” Journal of Robotics and Mechanical Engineering Research, vol. 1, no. 4, pp. 17–29, 2016. View at: Publisher Site | Google Scholar
  30. Y. Chen, S. Le, Q. C. Tan, O. Lau, F. Wan, and C. Song, “A lobster-inspired robotic glove for hand rehabilitation,” in 2017 IEEE International Conference on Robotics and Automation (ICRA), pp. 4782–4787, Singapore, Singapore, 2017. View at: Publisher Site | Google Scholar
  31. S. Biggar and W. Yao, “Design and Evaluation of a Soft and Wearable Robotic Glove for Hand Rehabilitation,” IEEE Transactions on Neural Systems and Rehabilitation Engineering, vol. 24, no. 10, pp. 1071–1080, 2016. View at: Publisher Site | Google Scholar
  32. J. Brackenridge, L. V. Bradnam, S. Lennon, J. J. Costi, and D. A. Hobbs, “A Review of Rehabilitation Devices to Promote Upper Limb Function Following Stroke,” Neuroscience and Biomedical Engineering, vol. 4, no. 1, pp. 25–42, 2016. View at: Publisher Site | Google Scholar
  33. G. A. Prieto, A. I. Andrés, A. Venkatakrishnan, W. Q. Malik, V. Dietz, and W. Z. Rymer, “Rehabilitation Technologies for Spinal Injury,” in Emerging Therapies in Neurorehabilitation II, pp. 65–85, Springer, 2016. View at: Google Scholar
  34. K. Serbest, S. Ateş, and A. H. A. Stienen, “Design of an exercise glove for hand rehabilitation using spring mechanism,” in 2016 20th National Biomedical Engineering Meeting (BIYOMUT), pp. 1–5, Izmir, Turkey, 2016. View at: Publisher Site | Google Scholar
  35. H. Cao and D. Zhang, “Soft robotic glove with integrated sEMG sensing for disabled people with hand paralysis,” in 2016 IEEE International Conference on Robotics and Biomimetics (ROBIO), pp. 714–718, Qingdao, China, 2016. View at: Publisher Site | Google Scholar
  36. A. Rodigari, A. Zancan, I. Jedrychowska, and I. Springhetti, “Effectiveness of the treatment of hemiplegic patient’s hand with robotic rehabilitation glove “gloreha”: preliminary results,” in The 42nd National Congress Of The Italian Society Of Physical And Rehabilitative Medicine, pp. 142–145, Torino, Italy, 2014, SIMFER. View at: Google Scholar
  37. P. Maeder-York, T. Clites, E. Boggs et al., “Biologically Inspired Soft Robot for Thumb Rehabilitation1,” Journal of Medical Devices, vol. 8, no. 2, 2014. View at: Publisher Site | Google Scholar
  38. P. Bernocchi, C. Mulè, F. Vanoglio, G. Taveggia, A. Luisa, and S. Scalvini, “Home-based hand rehabilitation with a robotic glove in hemiplegic patients after stroke: a pilot feasibility study,” Topics in Stroke Rehabilitation, vol. 25, pp. 114–119, 2018. View at: Publisher Site | Google Scholar
  39. J.-H. Shin, M.-Y. Kim, J.-Y. Lee et al., “Effects of virtual reality-based rehabilitation on distal upper extremity function and health-related quality of life: a single-blinded, randomized controlled trial,” Journal of NeuroEngineering and Rehabilitation, vol. 13, no. 1, 2016. View at: Publisher Site | Google Scholar
  40. B. Radder, G. B. Prange-Lasonder, A. I. R. Kottink, A. Melendez-Calderon, J. H. Buurke, and J. S. Rietman, “Feasibility of a wearable soft-robotic glove to support impaired hand function in stroke patients,” Journal of Rehabilitation Medicine, vol. 50, no. 7, pp. 598–606, 2018. View at: Publisher Site | Google Scholar
  41. H. K. Yap, J. H. Lim, F. Nasrallah, F. Z. Low, J. C. H. Goh, and R. C. H. Yeow, “MRC-glove: A fMRI compatible soft robotic glove for hand rehabilitation application,” in 2015 IEEE International Conference on Rehabilitation Robotics (ICORR), pp. 735–740, Singapore, Singapore, 2015. View at: Publisher Site | Google Scholar
  42. Y. Jiang, D. Chen, J. Que, Z. Liu, Z. Wang, and Y. Xu, “Soft robotic glove for hand rehabilitation based on a novel fabrication method,” in 2017 IEEE International Conference on Robotics and Biomimetics (ROBIO), pp. 817–822, Macau, China, 2017. View at: Publisher Site | Google Scholar
  43. H. K. Yap, B. W. K. Ang, J. H. Lim, J. C. H. Goh, and C. H. Yeow, “A fabric-regulated soft robotic glove with user intent detection using EMG and RFID for hand assistive application,” in 2016 IEEE International Conference on Robotics and Automation (ICRA), pp. 3537–3542, Stockholm, Sweden, 2016. View at: Publisher Site | Google Scholar
  44. H. K. Yap, J. H. Lim, F. Nasrallah, and C.-H. Yeow, “Design and Preliminary Feasibility Study of a Soft Robotic Glove for Hand Function Assistance in Stroke Survivors,” Frontiers in Neuroscience, vol. 11, 2017. View at: Publisher Site | Google Scholar
  45. F. Vanoglio, A. Luisa, F. Garofali, and C. Mora, “Evaluation of the effectiviness of Glohera (hand rehabilitation glove) on hemiplegic patients. Pilot study,” in XIII Congress of Italian Society of Neurorehabilitation, pp. 18–20, Bari, Italy, 2013. View at: Google Scholar
  46. G. B. Prange-Lasonder, B. Radder, A. I. R. Kottink, A. Melendez-Calderon, J. H. Buurke, and J. S. Rietman, “Applying a soft-robotic glove as assistive device and training tool with games to support hand function after stroke: Preliminary results on feasibility and potential clinical impact,” in 2017 International Conference on Rehabilitation Robotics (ICORR), pp. 1401–1406, London, UK, 2017. View at: Publisher Site | Google Scholar
  47. H. Kim, A. Lee, Y. I. Shin et al., “Effects of digital smart glove system on motor recovery of upper extremity in subacute stroke patients,” Annals of Physical and Rehabilitation Medicine, vol. 61, p. e28, 2018. View at: Publisher Site | Google Scholar
  48. M. Haghshenas-Jaryani, W. Carrigan, C. Nothnagle, and M. B. J. Wijesundara, “Sensorized soft robotic glove for continuous passive motion therapy,” in 2016 6th IEEE International Conference on Biomedical Robotics and Biomechatronics (BioRob), pp. 815–820, Singapore, Singapore, 2016. View at: Publisher Site | Google Scholar
  49. S. Toochinda and W. Wannasuphoprasit, “Design and Development of an Assistive Hand Device for Enhancing Compatibility and Comfortability,” in 2018 2nd International Conference on Engineering Innovation (ICEI), pp. 1–6, Bangkok, Thailand, 2018. View at: Publisher Site | Google Scholar
  50. P. Polygerinos, K. C. Galloway, S. Sanan, M. Herman, and C. J. Walsh, “EMG controlled soft robotic glove for assistance during activities of daily living,” in 2015 IEEE International Conference on Rehabilitation Robotics (ICORR), pp. 55–60, Singapore, Singapore, 2015. View at: Publisher Site | Google Scholar
  51. P. Polygerinos, Z. Wang, K. C. Galloway, R. J. Wood, and C. J. Walsh, “Soft robotic glove for combined assistance and at-home rehabilitation,” Robotics and Autonomous Systems, vol. 73, pp. 135–143, 2015. View at: Publisher Site | Google Scholar
  52. M. Xiloyannis, L. Cappello, B. Khanh Dinh, C. W. Antuvan, and L. Masia, “Design and Preliminary Testing of a Soft Exosuit for Assisting Elbow Movements and Hand Grasping,” in Converging Clinical and Engineering Research on Neurorehabilitation II, vol. 15, pp. 557–561, Springer, 2017. View at: Google Scholar
  53. M. Xiloyannis, L. Cappello, D. B. Khanh, S.-C. Yen, and L. Masia, “Modelling and design of a synergy-based actuator for a tendon-driven soft robotic glove,” in 2016 6th IEEE International Conference on Biomedical Robotics and Biomechatronics (BioRob), pp. 1213–1219, Singapore, Singapore, 2016. View at: Publisher Site | Google Scholar
  54. C. Jiménez, R. Mora, J. Pérez, and K. Quirós, “Nitiglove: Nitinol-Driven Robotic Glove Used to Assist Therapy for Hand Mobility Recovery,” in Wearable Robotics: Challenges and Trends, pp. 379–383, Springer, 2017. View at: Google Scholar
  55. A. Hadi, K. Alipour, S. Kazeminasab, and M. Elahinia, “ASR glove: A wearable glove for hand assistance and rehabilitation using shape memory alloys,” Journal of Intelligent Material Systems and Structures, vol. 29, no. 8, pp. 1575–1585, 2017. View at: Publisher Site | Google Scholar
  56. S. Sridar, P. H. Nguyen, M. Zhu, Q. P. Lam, and P. Polygerinos, “Development of a soft-inflatable exosuit for knee rehabilitation,” in 2017 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS), pp. 3722–3727, Vancouver, BC, Canada, 2017. View at: Publisher Site | Google Scholar
  57. H. Al-Fahaam, S. Davis, S. Nefti-Meziani, and T. Theodoridis, “Novel soft bending actuator-based power augmentation hand exoskeleton controlled by human intention,” Intelligent Service Robotics, vol. 11, no. 3, pp. 247–268, 2018. View at: Publisher Site | Google Scholar
  58. H. Al-Fahaam, S. Davis, and S. Nefti-Meziani, “The design and mathematical modelling of novel extensor bending pneumatic artificial muscles (EBPAMs) for soft exoskeletons,” Robotics and Autonomous Systems, vol. 99, pp. 63–74, 2018. View at: Publisher Site | Google Scholar
  59. M. Xiloyannis, L. Cappello, K. D. Binh, C. W. Antuvan, and L. Masia, “Preliminary design and control of a soft exosuit for assisting elbow movements and hand grasping in activities of daily living,” Journal of Rehabilitation and Assistive, vol. 4, article 205566831668031, 2017. View at: Publisher Site | Google Scholar
  60. H. Abidi and M. Cianchetti, “On intrinsic safety of soft robots,” Frontiers in Robotics and AI, vol. 4, no. 5, 2017. View at: Publisher Site | Google Scholar
  61. S. Jadhav, V. Kannanda, B. Kang, M. T. Tolley, and J. P. Schulze, “Soft robotic glove for kinesthetic haptic feedback in virtual reality environments,” Electronic Imaging, vol. 2017, no. 3, pp. 19–24, 2017. View at: Publisher Site | Google Scholar
  62. A. Avižienis, J. C. Laprie, B. Randell, and C. Landwehr, “Basic concepts and taxonomy of dependable and secure computing,” IEEE Transactions on Dependable and Secure Computing, vol. 1, no. 1, pp. 11–33, 2004. View at: Publisher Site | Google Scholar
  63. J. Guiochet, M. Machin, and H. Waeselynck, “Safety-critical advanced robots: a survey,” Robotics and Autonomous Systems, vol. 94, pp. 43–52, 2017. View at: Publisher Site | Google Scholar
  64. K. Chinzei, “Safety of surgical robots and IEC 80601-2-77: the first international standard for surgical robots,” Acta Polytechnica Hungarica, vol. 16, no. 8, pp. 171–184, 2019. View at: Publisher Site | Google Scholar
  65. Y. Tsuji, “A safety review of medical device robots in Japan,” sia Pacific Journal of Health Law &Ethics, vol. 12, no. 1, pp. 1–27, 2018. View at: Google Scholar
  66. P. Barattini, V. Federico, G. S. Virk, and T. Haidegger, Human–Robot Interaction, Taylor & F. Boca Raton: CRC Press, New York NY USA, 2019. View at: Publisher Site
  67. Z. Sun, Z. Guo, and W. Tang, “Design of wearable hand rehabilitation glove with soft hoop-reinforced pneumatic actuator,” Journal of Central South University, vol. 26, no. 1, pp. 106–119, 2019. View at: Publisher Site | Google Scholar
  68. H. Li and L. Cheng, “Preliminary study on the design and control of a pneumatically-actuated hand rehabilitation device,” in 2017 32nd Youth Academic Annual Conference of Chinese Association of Automation (YAC), pp. 860–865, Hefei, 2017. View at: Publisher Site | Google Scholar
  69. T. Jiralerspong, K. H. L. Heung, R. K. Y. Tong, and Z. Li, “A novel soft robotic glove for daily life assistance,” in 2018 7th IEEE International Conference on Biomedical Robotics and Biomechatronics (Biorob), pp. 671–676, Enschede, 2018. View at: Publisher Site | Google Scholar
  70. L. Wang, S. G. Nurzaman, and F. Iida, “Soft-material robotics,” Foundations and Trends in Robotics, vol. 5, no. 3, pp. 191–259, 2017. View at: Publisher Site | Google Scholar
  71. B. Wang, A. McDaid, M. Biglari-Abhari, and K. C. Aw, “Design and development of a glove for post-stroke hand rehabilitation,” in 2017 IEEE International Conference on Advanced Intelligent Mechatronics (AIM), pp. 1047–1051, Munich, 2017. View at: Publisher Site | Google Scholar
  72. D. H. Kim and H. S. Park, “Cable actuated dexterous (CADEX) glove for effective rehabilitation of the hand for patients with neurological diseases,” in 2018 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS), Madrid, Spain, 2018. View at: Publisher Site | Google Scholar
  73. M. A. Delph, S. A. Fischer, P. W. Gauthier, C. H. M. Luna, E. A. Clancy, and G. S. Fischer, “A soft robotic exomusculature glove with integrated sEMG sensing for hand rehabilitation,” in 2013 IEEE 13th International Conference on Rehabilitation Robotics (ICORR), pp. 1–7, Seattle, WA, 2013. View at: Publisher Site | Google Scholar
  74. A. Mohammadi, J. Lavranos, P. Choong, and D. Oetomo, “Flexo-glove: a 3D printed soft exoskeleton robotic glove for impaired hand rehabilitation and assistance,” in 2018 40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), pp. 2120–2123, Honolulu, HI, 2018. View at: Publisher Site | Google Scholar
  75. H. K. Yap, P. M. Khin, T. H. Koh et al., “A fully fabric-based bidirectional soft robotic glove for assistance and rehabilitation of hand impaired patients,” IEEE Robotics and Automation Letters, vol. 2, no. 3, pp. 1383–1390, 2017. View at: Publisher Site | Google Scholar
  76. H. K. Yap, J. C. H. Goh, and R. C. H. Yeow, “Design and characterization of soft actuator for hand rehabilitation application,” in 6th European Conference of the International Federation for Medical and Biological Engineering, Springer, pp. 367–370, Dubrovnik, 2015. View at: Publisher Site | Google Scholar
  77. H. K. Yap, J. H. Lim, F. Nasrallah, J. C. H. Goh, and C.-H. Yeow, “Characterisation and evaluation of soft elastomeric actuators for hand assistive and rehabilitation applications,” Journal of Medical Engineering & Technology, vol. 40, no. 4, pp. 199–209, 2016. View at: Publisher Site | Google Scholar
  78. H. K. Yap, F. Sebastian, C. Wiedeman, and C. H. Yeow, “Design and characterization of low-cost fabric-based flat pneumatic actuators for soft assistive glove application,” in 2017 International Conference on Rehabilitation Robotics (ICORR), pp. 1465–1470, London, 2017. View at: Publisher Site | Google Scholar
  79. A. F. Hassanin, D. Steve, and N. M. Samia, “A novel, soft, bending actuator for use in power assist and rehabilitation exoskeletons,” in 2017 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS), pp. 533–538, Vancouver, BC, 2017. View at: Publisher Site | Google Scholar
  80. P. Polygerinos, Z. Wang, J. T. B. Overvelde et al., “Modeling of soft fiber-reinforced bending actuators,” IEEE Transactions on Robotics, vol. 31, no. 3, pp. 778–789, 2015. View at: Publisher Site | Google Scholar
  81. B. Wang, K. C. Aw, M. Biglari-Abhari, and A. McDaid, “Design and fabrication of a fiber-reinforced pneumatic bending actuator,” in 2016 IEEE International Conference on Advanced Intelligent Mechatronics (AIM), pp. 83–88, Banff, 2016. View at: Publisher Site | Google Scholar
  82. Y. Jiang, D. Chen, P. Liu et al., “Fishbone-inspired soft robotic glove for hand rehabilitation with multi-degrees-of-freedom,” in 2018 IEEE International Conference on Soft Robotics (RoboSoft), pp. 394–399, Livorno, 2018. View at: Publisher Site | Google Scholar
  83. H. Lee, B. B. Kang, H. In, and K.-J. Cho, Design improvement of a polymer-based tendon-driven wearable robotic hand (exo-glove poly), vol. 16 of Wearable Robotics: Challenges and Trends, no. 1, Springer, 2017. View at: Publisher Site
  84. M. Nilsson, J. Ingvast, J. Wikander, and H. von Holst, “The soft extra muscle system for improving the grasping capability in neurological rehabilitation,” in 2012 IEEE-EMBS Conference on Biomedical Engineering and Sciences, pp. 412–417, Langkawi, 2012. View at: Publisher Site | Google Scholar
  85. F. L. Hammond, Y. Mengüç, and R. J. Wood, “Toward a modular soft sensor-embedded glove for human hand motion and tactile pressure measurement,” in 2014 IEEE/RSJ International Conference on Intelligent Robots and Systems, pp. 4000–4007, Chicago, IL, 2014. View at: Publisher Site | Google Scholar
  86. F. Schmitt, O. Piccin, L. Barbé, and B. Bayle, “Soft robots manufacturing: a review,” Frontiers in Robotics and AI, vol. 5, pp. 84–101, 2018. View at: Publisher Site | Google Scholar
  87. P. Polygerinos, N. Correll, S. A. Morin et al., “Soft Robotics: Review of Fluid‐Driven Intrinsically Soft Devices; Manufacturing, Sensing, Control, and Applications in Human‐Robot Interaction,” Advanced Engineering Materials, vol. 19, no. 12, pp. 1–22, 2017. View at: Publisher Site | Google Scholar
  88. M. Haghshenas-Jaryani, C. Nothnagle, R. M. Patterson, N. Bugnariu, and M. B. J. Wijesundara, “Soft robotic rehabilitation exoskeleton (rehab glove) for hand therapy,” in Volume 3: 19th International Conference on Advanced Vehicle Technologies; 14th International Conference on Design Education; 10th Frontiers in Biomedical Devices, Cleveland, OH, 2017. View at: Publisher Site | Google Scholar
  89. B. W. K. Ang and C. H. Yeow, “Print-it-Yourself (PIY) glove: a fully 3D printed soft robotic hand rehabilitative and assistive exoskeleton for stroke patients,” in 2017 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS), pp. 1219–1223, Vancouver, 2017. View at: Publisher Site | Google Scholar
  90. C. G. Rose and M. K. O'Malley, “Hybrid rigid-soft hand exoskeleton to assist functional dexterity,” IEEE Robotics and Automation Letters, vol. 4, no. 1, pp. 73–80, 2019. View at: Publisher Site | Google Scholar
  91. M. Li, B. He, Z. Liang et al., “An attention-controlled hand exoskeleton for the rehabilitation of finger extension and flexion using a rigid-soft combined mechanism,” Frontiers in Neurorobotics, vol. 13, no. 34, pp. 1–13, 2019. View at: Publisher Site | Google Scholar
  92. B. Radder, A. I. R. Kottink, N. van der Vaart et al., “User-centred input for a wearable soft-robotic glove supporting hand function in daily life,” in 2015 IEEE International Conference on Rehabilitation Robotics (ICORR), pp. 502–507, Singapore, 2015. View at: Publisher Site | Google Scholar
  93. C. D. Takahashi, L. Der-Yeghiaian, V. Le, R. R. Motiwala, and S. C. Cramer, “Robot-based hand motor therapy after stroke,” Brain, vol. 131, no. 2, pp. 425–437, 2008. View at: Publisher Site | Google Scholar
  94. T. Proietti, V. Crocher, A. Roby-Brami, and N. Jarrassé, “Upper-limb robotic exoskeletons for neurorehabilitation: a review on control strategies,” IEEE Reviews in Biomedical Engineering, vol. 9, pp. 4–14, 2016. View at: Publisher Site | Google Scholar
  95. C. M. L. Hughes, P. Tommasino, A. Budhota, and D. Campolo, “Upper extremity proprioception in healthy aging and stroke populations, and the effects of therapist- and robot-based rehabilitation therapies on proprioceptive function,” Frontiers in Human Neuroscience, vol. 9, 2015. View at: Publisher Site | Google Scholar
  96. M. K. Burns, D. Pei, and R. Vinjamuri, “Myoelectric control of a soft hand exoskeleton using kinematic synergies,” Tech. Rep. 6, IEEE Transactions on Biomedical Circuits and Systems, 2019. View at: Publisher Site | Google Scholar
  97. H. K. Yap, A. Mao, J. C. H. Goh, and C. H. Yeow, “Design of a wearable FMG sensing system for user intent detection during hand rehabilitation with a soft robotic glove,” in 2016 6th IEEE International Conference on Biomedical Robotics and Biomechatronics (BioRob), pp. 781–786, Singapore, 2016. View at: Publisher Site | Google Scholar
  98. N. F. M. Nawi, A. H. K. Ching, N. Ibrahim et al., “Design and development approach of smart glove for post stroke rehabilitation,” International Journal of Engineering & Technology, vol. 7, no. 4, pp. 78–82, 2018. View at: Google Scholar
  99. S. Sasidharan, Smart glove: an assistive device to enhanced recovery of hand function during motor rehabilitation, [M.S. thesis], Arizona State University, 2015.
  100. G. Rudd, L. Daly, V. Jovanovic, and F. Cuckov, “A low-cost soft robotic hand exoskeleton for use in therapy of limited hand-motor function,” Tech. Rep. 18, Applied Sciences, 2019. View at: Publisher Site | Google Scholar
  101. M. Haghshenas-Jaryani, C. Pande, and B. J. Muthu Wijesundara, “Soft robotic bilateral hand rehabilitation system for fine motor learning,” pp. 337–342, Toronto, ON, 2019. View at: Publisher Site | Google Scholar
  102. T. George Thuruthel, Y. Ansari, E. Falotico, and C. Laschi, “Control strategies for soft robotic manipulators: a survey,” Soft Robotics, vol. 5, no. 2, pp. 149–163, 2018. View at: Publisher Site | Google Scholar
  103. D. Popescu, M. Ivanescu, R. Popescu, A. Petrisor, L. C. Popescu, and A. M. Bumbea, “Post-stroke hand rehabilitation using a wearable robotic glove,” in Smart Innovation, Systems and Technologies, pp. 259–268, Tenerife, Spain, 2016. View at: Publisher Site | Google Scholar
  104. J. Nassour and F. Hamker, “Enfolded textile actuator for soft wearable robots,” in 2019 IEEE International Conference on Cyborg and Bionic Systems (CBS), pp. 60–65, Munich, Germany, 2019. View at: Publisher Site | Google Scholar
  105. F. Vanoglio, P. Bernocchi, C. Mulè et al., “Feasibility and efficacy of a robotic device for hand rehabilitation in hemiplegic stroke patients: a randomized pilot controlled study,” Clinical Rehabilitation, vol. 31, no. 3, pp. 351–360, 2017. View at: Publisher Site | Google Scholar
  106. H. K. Yap, N. Kamaldin, J. H. Lim, F. A. Nasrallah, J. C. H. Goh, and C. H. Yeow, “A magnetic resonance compatible soft wearable robotic glove for hand rehabilitation and brain imaging,” IEEE Transactions on Neural Systems and Rehabilitation Engineering, vol. 25, no. 6, pp. 782–793, 2017. View at: Publisher Site | Google Scholar
  107. D. Popov, I. Gaponov, and J. H. Ryu, “Portable exoskeleton glove with soft structure for hand assistance in activities of daily living,” IEEE/ASME Transactions on Mechatronics, vol. 22, no. 2, pp. 865–875, 2017. View at: Publisher Site | Google Scholar
  108. K. Shiota, S. Kokubu, T. V. J. Tarvainen et al., “Enhanced Kapandji test evaluation of a soft robotic thumb rehabilitation device by developing a fiber-reinforced elastomer-actuator based 5-digit assist system,” Robotics and Autonomous Systems, vol. 111, pp. 20–30, 2019. View at: Publisher Site | Google Scholar
  109. B. Kim, H. In, D. Y. Lee, and K. J. Cho, “Development and assessment of a hand assist device: GRIPIT,” Journal of Neuroengineering and Rehabilitation, vol. 14, no. 1, 2017. View at: Publisher Site | Google Scholar
  110. A. L. Coffey, D. J. Leamy, and T. E. Ward, “A novel BCI-controlled pneumatic glove system for home-based neurorehabilitation,” in 2014 36th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, pp. 3622–3625, Chicago, IL, 2014. View at: Publisher Site | Google Scholar
  111. A. M. Gordon, Y. C. Hung, M. Brandao et al., “Bimanual training and constraint-induced movement therapy in children with hemiplegic cerebral palsy,” Neurorehabilitation and Neural Repair, vol. 25, no. 8, pp. 692–702, 2011. View at: Publisher Site | Google Scholar
  112. L. Gerez, J. Chen, and M. Liarokapis, “On the development of adaptive, tendon-driven, wearable exo-gloves for grasping capabilities enhancement,” IEEE Robotics and Automation Letters, vol. 4, no. 2, pp. 422–429, 2019. View at: Publisher Site | Google Scholar
  113. Glohera, “Hand rehabilitation glove,” 2019, http://www.gloreha.com/?lang=it. View at: Google Scholar
  114. D. Ryu, K.-W. Moon, H. Nam et al., “Micro hydraulic system using slim artificial muscles for a wearable haptic glove,” in 2008 IEEE/RSJ International Conference on Intelligent Robots and Systems, pp. 3028–3033, Nice, 2008. View at: Publisher Site | Google Scholar
  115. Neofect, “Hand assistance glove,” 2019, http://www.neofect.com/en/product/rapael/. View at: Google Scholar
  116. J. B. Chossat, Y. Tao, V. Duchaine, and Y. L. Park, “Wearable soft artificial skin for hand motion detection with embedded microfluidic strain sensing,” in 2015 IEEE International Conference on Robotics and Automation (ICRA), pp. 2568–2573, Seattle, Washington, 2015. View at: Publisher Site | Google Scholar
  117. B. Radder, G. B. Prange-Lasonder, A. I. R. Kottink et al., “A wearable soft-robotic glove enables hand support in ADL and rehabilitation: a feasibility study on the assistive functionality,” Journal of Rehabilitation and Assistive Technologies Engineering, vol. 3, 2016. View at: Publisher Site | Google Scholar
  118. B. Radder, G. B. Prange-Lasonder, A. I. R. Kottink et al., “The effect of a wearable soft-robotic glove on motor function and functional performance of older adults,” Assistive Technology, vol. 32, pp. 1–27, 2020. View at: Publisher Site | Google Scholar
  119. J. Yi, Z. Shen, C. Song, and Z. Wang, “A soft robotic glove for hand motion assistance,” in IEEE International Conference on Real-time Computing and Robotics (RCAR), pp. 111–116, Angkor, 2016. View at: Publisher Site | Google Scholar
  120. J. Yi, X. Chen, and Z. Wang, “A three-dimensional-printed soft robotic glove with enhanced ergonomics and force capability,” IEEE Robotics and Automation Letters, vol. 3, no. 1, pp. 242–248, 2018. View at: Publisher Site | Google Scholar
  121. Z. Yao, C. Linnenberg, A. Argubi-Wollesen, R. Weidner, and J. P. Wulfsberg, “Biomimetic design of an ultra-compact and light-weight soft muscle glove,” Production Engineering, vol. 11, no. 6, pp. 731–743, 2017. View at: Publisher Site | Google Scholar
  122. S. Chatterjee, D. Hore, and A. Arora, “A wearable soft pneumatic finger glove with antagonistic actuators for finger rehabilitation,” in 2017 2nd International Conference on Communication and Electronics Systems (ICCES), pp. 341–345, Coimbatore, India, 2017. View at: Publisher Site | Google Scholar
  123. K. O. Thielbar, K. M. Triandafilou, H. C. Fischer et al., “Benefits of using a voice and EMG-driven actuated glove to support occupational therapy for stroke survivors,” IEEE Transactions on Neural Systems and Rehabilitation Engineering, vol. 25, no. 3, pp. 297–305, 2017. View at: Publisher Site | Google Scholar
  124. K. H. L. Heung, R. K. Y. Tong, A. T. H. Lau, and Z. Li, “Robotic glove with soft-elastic composite actuators for assisting activities of daily living,” Soft Robotics, vol. 6, no. 2, pp. 289–304, 2019. View at: Publisher Site | Google Scholar
  125. L. Cappello, J. T. Meyer, K. C. Galloway et al., “Assisting hand function after spinal cord injury with a fabric-based soft robotic glove,” Journal of NeuroEngineering and Rehabilitation, vol. 15, no. 1, 2018. View at: Publisher Site | Google Scholar
  126. P. I. Daiya, “Daiya glove,” 2015, https://exoskeletonreport.com/product/daiya-glove/. View at: Google Scholar
  127. K. Toya, T. Miyagawa, and Y. Kubota, “Power-assist glove operated by predicting the grasping mode,” Journal of System Design and Dynamics, vol. 5, no. 1, pp. 94–108, 2011. View at: Publisher Site | Google Scholar
  128. C.-Y. Chu and R. M. Patterson, “Soft robotic devices for hand rehabilitation and assistance: a narrative review,” Journal of NeuroEngineering and Rehabilitation, vol. 15, no. 1, pp. 9–14, 2018. View at: Publisher Site | Google Scholar
  129. Z. Ouyang, P. Zhou, J. Du, and H. Chi, “Modified technology readiness level for advanced medical devices in China,” Western Pacific Region Index Medicus, vol. 28, no. 3, pp. 217–221, 2015. View at: Google Scholar
  130. ISO, “ISO 16290:2013,” 2013, https://www.iso.org/standard/56064.html. View at: Google Scholar

Copyright © 2020 Juana-Mariel Dávila-Vilchis et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.


More related articles

 PDF Download Citation Citation
 Download other formatsMore
 Order printed copiesOrder
Views218
Downloads129
Citations

Related articles

We are committed to sharing findings related to COVID-19 as quickly as possible. We will be providing unlimited waivers of publication charges for accepted research articles as well as case reports and case series related to COVID-19. Review articles are excluded from this waiver policy. Sign up here as a reviewer to help fast-track new submissions.