Correlation obtained between sonoclot parameters and conventional coagulation variables in patients with liver disease.

Conventional testsPT-INRAPTTFibrinogenD-dimerPlatelet count

Sonoclot parameters
value ( : correlation coefficient)
SonACT
<0.008 ( )<0.0015 ( )<0.037 ( )0.39 ( )<0.01 ( )
CR<0.0025 ( )<0.0025 ( )<0.004 ( )0.96 ( )<0.03 ( )
PF0.07 ( )0.2 ( )<0.001 ( )0.89 ( )<0.0001 ( )
TP<0.0002 ( )<0.01 ( )<0.0001 ( )0.12 ( )<0.0001 ( )
PA<0.0006 ( )<0.008 ( )<0.0001 ( )0.6 ( )<0.0002 ( )

PT: prothrombin time, INR: international normalized ratio, APTT: activated partial thromboplastin time.
SONACT: sonoclot activated clotting time, CR: clot rate, PF: platelet function, TP: time to peak, PA: peak amplitude.
: Spearmans rank correlation coefficient.