Assessment of Ground-Based and Aerial Cloud Seeding Using Trace Chemistry
Table 1
Operating ICP-MS conditions and acquisition parameters.
| Parameter | Setting value |
| Forward power | 1400 W | Cooling | 13 L/min | Expansion pressure | 2.0 100 mbar | Analyzer pressure | 4.0–4.2 10−7 mbar | Turbo pump speed | 1000 Hz | Vacuum pump load | 0.8–1.2 A | Pole bias | 1.0 | Hex bias | 4.0 | Sample cone | Ni, 1.1 mm aperture inner diameter | Microskimmer cone | Ni, 0.8 mm aperture inner diameter | Sample depth | Adjusted daily | Lens settings | Adjusted when appropriate | Nebulizer back pressure | ~2.0 bar, optimized daily | Nebulizer Flow Rate | 400 μL/min | Sample Uptake Time | 20 s | Rinse Time between samples | 20 s | Elements measured | Na, Al, Cr, Co, Sr, Ag, Ba, La, Ce, Pb | Dwell time per element | Ag: 200 ms; Na, Al, Cr, Co, Sr, Ba, La, Ce, Pb: 10 ms |
|
|
Optimized to obtain a stable/high 115 In signal (>500,000 cps for 10 ppb) min. oxide formation rate. |