Research Article

Comparison of the Tendon Damage Caused by Four Different Anchor Systems Used in Transtendon Rotator Cuff Repair

Table 1

Areas of tendon damage caused by anchors.

Sample number(a) Healix Peek anchor (5.5 mm)(b) Fastin RC anchor (5.0 mm)(c) Bio-Corkscrew Suture anchor (5.0 mm)(d) Healix Transtend anchor inserted through the Percannula system (4.0 mm)(e) Healix Transtend anchor (3.4 mm) without cannula

125.017.423.37.013.0
228.622.621.410.710.0
331.622.822.99.110.1
437.518.624.710.916.7
525.822.722.96.618.7
626.818.323.77.514.1
728.520.224.811.713.1
Mean29.120.423.49.113.7
SD4.32.31.22.13.2
(versus a) 0.00050.00520.00000.0000
(versus b) 0.00990.00000.0008
(versus c) 0.00000.0000
(versus d) 0.0078

Note: SD represents standard deviation. values were obtained by the two-tailed Student’s t-test, comparing between two types of anchors as indicated.