Research Article

Development and Validation of the Quantification Method for Hydroxychloroquine in Volumetric Absorptive Microsampling (VAMS) Using High-Performance Liquid Chromatography-Photodiode Array

Table 1

Data of the system suitability test.

ReplicateRetention timeArea (µV/s)
HCQCQHCQCQPAR

15.289.902803563438420.81536
25.299.902858793463770.82534
35.289.912822293407340.82830
45.299.902805043376270.83081
55.299.912785383329180.83666
65.309.922737153332430.82137
Mean5.299.91280203.50339123.500.82631
SD0.010.014032.425528.100.00744
CV (%)0.110.091.441.630.90