Development and Validation of the Quantification Method for Hydroxychloroquine in Volumetric Absorptive Microsampling (VAMS) Using High-Performance Liquid Chromatography-Photodiode Array

Table 3

Data of accuracy and precision, within-run, and between-run.

Nominal concentration (ng/mL)DayWithin-runBetween-run
Mean measured concentration (ng/mL)CV (%)% difference (range)Mean measured concentration (ng/mL)CV (%)

LLOQ (2.00)11.9710.07−17.01–7.691.979.24
21.9811.42−14.20–16.67
31.978.17−11.91–8.79

QCL (6.00)15.716.15−9.67–5.925.927.15
25.918.30−11.19–7.42
36.156.12−3.66–10.11

QCM (2,600)12632.537.19−5.15–11.502605.698.12
22434.376.92−12.02–3.57
32750.186.17−0.98–14.06

QCH (4,785)15110.595.97−4.45–12.954903.429.46
24504.877.82−12.86–4.78
35094.819.58−12.19–13.18