Leakage Characteristics of Dual-Cannula Fenestrated Tracheostomy Tubes during Positive Pressure Ventilation: A Bench Study

Figure 6

Loss of delivered minute volume of Portex Blue Line Ultra (+), TRACOE twist (×), or Rüsch Traceofix () tracheostomy tubes plotted against transfenestration leakage rate measured at PEEP settings of 3, 6, 9, and 12 mbar. Measurements were obtained in volume-controlled ventilation (ventilator rate: 15 breaths·min−1, tidal volume: 450 mL) at °C. Lung simulator compliance: 20 mL·mbar−1. TFLR: transfenestration leakage rate; : inspiratory minute ventilation; : expiratory minute ventilation.