Anesthesiology Research and Practice
Volume 2017 (2017), Article ID 2032748, 7 pages
https://doi.org/10.1155/2017/2032748
Clinical Study

Achieving the Recommended Endotracheal Tube Cuff Pressure: A Randomized Control Study Comparing Loss of Resistance Syringe to Pilot Balloon Palpation

1Department of Anaesthesia and Critical Care, Faculty of Health Sciences, Busitema University, Mbale, Uganda
2Department of Anaesthesia and Critical Care, College of Health Sciences, Makerere University, Kampala, Uganda

Correspondence should be addressed to Fred Bulamba; moc.liamg@abmalubylfderf

Received 27 June 2017; Accepted 24 October 2017; Published 31 December 2017

Academic Editor: Ferenc Petak

Copyright © 2017 Fred Bulamba et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

Background. Both under- and overinflation of endotracheal tube cuffs can result in significant harm to the patient. The optimal technique for establishing and maintaining safe cuff pressures (20–30 cmH2O) is the cuff pressure manometer, but this is not widely available, especially in resource-limited settings where its use is limited by cost of acquisition and maintenance. Therefore, anesthesia providers commonly rely on subjective methods to estimate safe endotracheal cuff pressure. This study set out to determine the efficacy of the loss of resistance syringe method at estimating endotracheal cuff pressures. Methods. This was a randomized clinical trial. We enrolled adult patients scheduled to undergo general anesthesia for elective surgery at Mulago Hospital, Uganda. Study participants were randomized to have their endotracheal cuff pressures estimated by either loss of resistance syringe or pilot balloon palpation. The pressures measured were recorded. Results. One hundred seventy-eight patients were analyzed. 66.3% (59/89) of patients in the loss of resistance group had cuff pressures in the recommended range compared with 22.5% (20/89) from the pilot balloon palpation method. This was statistically significant. Conclusion. The loss of resistance syringe method was superior to pilot balloon palpation at administering pressures in the recommended range. This method provides a viable option to cuff inflation.