Fluid temperature before and after warming.

Infusion rate (mL/h)Fluid temperature before warming (°C)Fluid temperature after warming (°C)Difference in fluid temperature (°C) value

10025.0 ± 0.135.9 ± 0.110.9 ± 0.1<0.001
30024.8 ± 0.136.3 ± 0.211.5 ± 0.1<0.001
60025.0 ± 0.235.2 ± 0.110.2 ± 0.1<0.001
90024.1 ± 0.734.2 ± 0.010.1 ± 0.7<0.001
1,20024.5 ± 0.233.0 ± 0.28.4 ± 0.2<0.001

Data are presented as mean ± standard deviation. Statistically significant at .