Research Article

Dimensional Variations of Left-Sided Double-Lumen Endobronchial Tubes

Table 4

Measured length of investigated left-sided double-lumen endobronchial tubes (DLTs).

ParameterSize (Fr)Sheridan DLTMallinckrodt DLTPortex DLTRüsch DLTVIVASIGHT DLTWell Lead DLT

A–C35260.0 (254.0–261.0)253.0 (252.0–254.0)318.5 (315.0–322.0)210.0 (207.0–214.0; 66%)241.0 (235.0–247.0)250.0 (248.0–253.0)
37258.0 (256.0–261.0)250.5 (249.0–252.0)317.5 (316.0–319.0)215.0 (214.0–215.0; 68%)241.0 (241.0–243.0)252.5 (248.0–256.0)
39263.0 (262.0–263.0)248.0 (247.0–248.5)307.5 (304.0–308.0)234.5 (233.0–236.0; 76%)241.0 (238.0–243.0)251.0 (246.0–256.0)
41264.5 (264.0–265.0)249.5 (247.0–251.0)304.5 (301.0–308.0)235.0 (234.0–236.0; 77%)242.0 (239.0–246.0)243.0 (239.0–247.0)

C-D3537.3 (36.0–40.0)40.0 (40.0–40.0)32.0 (32.0–32.8; 73%)36.3 (36.0–37.0)44.0 (43.7–44.3)44.0 (43.0–45.0)
3736.8 (35.5–38.0; 86%)43.0 (42.0–43.5)38.1 (37.0–39.0)37.0 (35.0–39.5)41.9 (41.7–43.0)43.0 (42.0–44.0)
3938.3 (38.0–39.0)43.0 (42.0–44.0)39.0 (37.0–42.0)37.5 (36.5–39.0; 83%)45.2 (45.0–45.5)42.0 (41.0–42.5)
4138.0 (37.0–40.0; 86%)39.0 (38.0–39.1)42.9 (41.0–44.0)38.1 (38.0–39.0)44.2 (43.3–46.0)42.3 (42.0–43.0)

D-E3538.0 (36.0–40.0)35.0 (35.0–35.0)40.0 (39.0–41.0)34.5 (31.0–38.0:78%)36.5 (35.6–37.0)44.5 (39.0–48.0)
3743.5 (42.0–45.0)34.3 (34.0–35.0; 79%)39.5 (37.0–42.0)40.5 (39.0–41.0)40.8 (40.0–41.0)42.5 (38.0–46.0)
3948.0 (45.0–51.0)40.0 (38.5–42.0)38.5 (35.0–41.0; 80%)42.0 (40.0–43.0)39.0 (37.0–40.3)44.5 (40.0–48.0)
4142.0 (38.0–45.0)47.0 (44.0–49.0)39.0 (36.0–42.0; 76%)52.0 (51.0–53.0)40.3 (39.0–41.8)51.5 (51.0–52.0)

E-F3523.0 (23.0–23.5)16.8 (16.5–17.0)24.5 (23.0–25.3)17.4 (16.0–18.3)14.9 (14.3–15.0; 61%)15.7 (15.4–16.1)
3722.5 (22.0–23.0)19.5 (19.0–20.0)25.4 (24.8–27.0)16.6 (16.0–18.0)17.1 (17.0–17.6)15.9 (15.1–16.5; 63%)
3918.0 (18.0–18.3)18.5 (17.5–19.0)30.0 (30.0–30.3)19.0 (18.0–19.0)17.3 (16.0–17.6)15.8 (15.0–16.0; 53%)
4121.5 (21.0–22.0)17.1 (17.0–18.0)30.8 (30.0–31.0)16.5 (16.3–17.0)17.1 (17.0–17.5)15.6 (15.0–16.1; 51%)

F-G358.1 (7.0–9.5)11.1 (10.9–11.3)4.6 (4.0–5.7; 28%)6.0 (5.0–6.5)16.4 (15.5–17.0)11.2 (10.3–11.5)
378.0 (7.3–8.1)12.0 (11.5–12.1)4.8 (4.5–4.8; 32%)7.1 (6.0–7.6)14.9 (12.5–18.0)12.0 (11.9–12.7)
3910.0 (9.5–10.0)11.7 (11.1–11.9)6.6 (6.3–6.8)6.0 (5.0–6.1; 46%)13.1 (12.9–14.0)10.6 (10.5–11.0)
418.6 (8.0–9.0)14.2 (14.0–14.4)7.0 (6.8–7.5; 48%)7.5 (7.0–8.0)14.6 (14.0–15.0)10.5 (10.0–11.7)

A–G35366.4 (356.0–374.0)355.9 (354.4–357.3)419.6 (413.0–426.7)304.1 (295.0–313.8; 72%)352.7 (344.1–360.3)365.4 (355.6–373.6)
37368.8 (362.8–375.1)359.3 (355.5–362.6)425.3 (419.3–431.8)316.2 (310.0–321.1; 74%)355.6 (352.1–362.6)365.9 (355.0–375.2)
39377.3 (372.5–381.3)361.2 (356.1–365.4)421.6 (412.3–428.0)339.0 (332.5–343.1; 80%)355.5 (348.9–360.4)363.8 (352.5–373.5)
41374.6 (368.0–381.0)366.8 (360.0–371.5)424.1 (414.8–432.5)349.1 (346.3–353.0; 82%)358.2 (352.3–366.3)362.8 (357.0–369.8)

A–C = distance from bifurcation (beginning of DLT shaft, A) to the proximal edge of the tracheal cuff (C); C-D = length of the tracheal cuff; D-E = distance between distal edge of the tracheal cuff (D) to the proximal edge of the bronchial Cuffs (E); E-F = length of the bronchial cuff; F-G = distance between the distal edge of the bronchial Cuffs (F) to the tip of the tube (G); A–G = tube length from bifurcation (A) to the tip of the tube (G) (see also Figure 1). DLT sizes are in French (= 1/3 mm). Data are displayed as median (minimum–maximum) given as millimeter (mm). The greatest length of a respective DLT size is shaded in gray (100%), and the smallest diameter is marked with an asterisk () (proportion of largest OD).