Research Article

A New Medical Device to Provide Independent Ventilation to Two Subjects Using a Single Ventilator: Evaluation in Lung-Healthy Pigs

Figure 3

Two animals, weighing 12 kg (left) and 25 kg (right), are shown being ventilated simultaneously with a single ventilator and the DuplicARⓇ device. The pressure-time curve of each animal is displayed in cmH2O in each multiparameter monitor (in detail in (a) and (b)). In (a), the peak pressure is 20.4 cmH2O, while the PEEP is 1.4 cmH20. In (b), the peak pressure is 23.1 cmH20, and the PEEP is 10.9 cmH20.