Combined Ventilation of Two Subjects with a Single Mechanical Ventilator Using a New Medical Device: An In Vitro Study

Table 1

Summary of the three stages, simulating the different possible combinations of subjects’ weights and lung compliances. The ventilator is configured in each mode to adequately ventilate the two subjects with a target volume of 7 ml/kg. The controllers of the DuplicAR device are manipulated to deliver asymmetric flow or different PEEP to the subjects.

Stage 1Stage 2Stage 3
ABABAB

Weight (kg)505050505028
Compliance (ml/cm H2O)3030203010
Compliance (ml cm H2O−1 kg−1)0.60.60.40.60.35
PEEP ventilator (cm H2O)555
I : E ratio1 : 21 : 21 : 2
Target volume (ml/kg)777
Total target load (ml)350350350350350196
Peak pressure in PCV (cm H2O)152122
Tidal volume in VCV (ml)700700546
DuplicAR flow manipulationNoNoYesNoYesNo
PEEP experiment I (cm H2O)555
PEEP experiment II (cm H2O)510510510
PEEP experiment III (cm H2O)515515515