A New Liquid Chromatography-Tandem Mass Spectrometry Method for Determination of Bisoprolol in Human Plasma Samples
Table 3
Evaluation of interday precision and accuracy for bisoprolol spiked quality control samples.
| = 3 ng/mL | = 25 ng/mL | = 75 ng/mL | | (ng/mL) | Accuracy (%) | (ng/mL) | Accuracy (%) | (ng/mL) | Accuracy (%) |
| (1) | 3.015 | 101.860 | 24.331 | 98.586 | 70.921 | 95.788 | (2) | 2.879 | 97.276 | 23.332 | 94.539 | 77.073 | 104.097 | (3) | 3.257 | 110.048 | 23.109 | 93.636 | 77.932 | 105.256 | (4) | 2.964 | 100.136 | 24.013 | 97.297 | 70.590 | 95.340 | (5) | 2.943 | 99.439 | 28.411 | 115.117 | 75.582 | 102.083 | (6) | 3.152 | 106.480 | 24.910 | 100.933 | 74.865 | 101.115 | Mean | 3.035 | 102.540 | 24.684 | 100.018 | 74.494 | 100.613 | SD | 0.142 | | 1.940 | | 3.092 | | CV % | 4.686 | | 7.860 | | 4.151 | |
| | | Acceptance criteria | | |
| | | 67% Total QCs must be % nominal values | | | | | 50% QCs per level must be % nominal values | | | | | Mean % nominal % | | | | | CV (%) 15% | | |
|
|