An Assessment of Hazards Caused by Electromagnetic Interaction on Humans Present near Short-Wave Physiotherapeutic Devices of Various Types Including Hazards for Users of Electronic Active Implantable Medical Devices (AIMD)

Table 1

The main parameters of investigated short-wave diathermia (SWD).

SWD typeTechnical parameters of SWD Settings of technical parameters
during measurements
The type of applicatorThe averaged output power regulation range (W)Duty cycle regulation range, Averaged output power (W)Duty cycle,

Phyaction Performa+Thermoplode,
Ø 14 cm/H
4.5–900.0016–0.45900.45

Curapuls 419Circuplode,
Ø 14 cm/H
6.3–840.0063–0.084840.084

Thermo 500Thermoplode,
Ø 14 cm/H
5–700.0016–0.35700.35

Curapuls 670Circuplode-E,
 cm/H
0.34–320.0017–0.16320.16

Curapuls 670Circuplode,
Ø 9 cm/H
0.17–160.0017–0.16160.16

Curapuls 670Circuplode,
Ø 9 cm/H
0.34–320.0017–0.16320.16

Terapuls GS 200Ø 20 cm/H4.8–600.0048–0.06600.06

Phyaction Performa+Thermoplode,
Ø 8 cm/H
3.2–320.0016–0.32320.32

Thermatur 500Circular plates,
Ø 16 cm/E
9–90Continuous wave (CW)80CW

Cosmogamma SW500Circular plates, Ø 14 cm/E0–500Continuous wave (CW)500CW

ThermopulseCircular plates,
Ø 15 cm/E
10–150Continuous wave (CW)150CW

The type of applicators: H: inductive, E: capacitive.