Total Body Irradiation with Step Translation and Dynamic Field Matching

Table 1

The optimal overlap widths of the two dynamic field edges matched at a distance of 220 cm from the radiation source. When overlapped optimally, the dose variation is within ±10% at the field junction.

Leaf motion length (cm)2345

Optimal overlap width (cm)3.5 5.0–5.56.5–7.58.5–10.0