Evaluation of Wall Correction Factor of INER's Air-Kerma Primary Standard Chamber and Dose Variation by Source Displacement for HDR 192Ir Brachytherapy

Table 2

Comparison of values calculated by INER and NIST for NIST’s spherical graphite primary standard chambers using 0.40 MeV, 0.662 MeV, and 1.25 MeV photons.

Chamber (0.40 MeV) (0.662 MeV) (1.25 MeV)
NISTINERDifferenceNISTINERDifferenceNISTINERDifference

1cc1.03121.03200.07%1.02861.02870.01%1.01971.02870.01%
10cc1.03491.03520.03%1.03141.03190.05%1.02261.03190.05%
30cc1.03741.03950.20%1.03481.0344−0.04%1.02491.0344−0.04%
50cc-11.03861.04020.16%1.03491.03540.05%1.02521.03540.05%