Evaluating the Therapeutic Dose Distribution of Intensity-Modulated Radiation Therapy for Head and Neck with Cone-Beam Computed Tomography Image: A Methodological Study

Table 1

Gamma pass rate of dose comparison between CBCT- and FBCT-based head phantom hybrid plans for 10 head and neck IMRT cases.

Case numberTransversal SagittalCoronal
3%—3 mm2%—2 mm3%—3 mm2%—2 mm3%—3 mm2%—2 mm

199.75%97.22%99.46%97.49%99.50%98.53%
299.94%99.27%99.57%97.47%99.56%98.98%
399.86%98.83%99.29%96.93%99.27%98.54%
499.89%98.77%99.38%97.37%99.26%98.57%
599.84%98.77%99.36%97.13%99.31%98.47%
699.89%98.79%100.00%99.94%100.00%99.54%
799.84%98.88%99.44%97.45%99.45%98.70%
899.88%98.96%99.21%97.03%99.18%98.35%
999.85%98.60%99.64%98.00%99.98%99.61%
1099.84%98.74%99.44%97.22%99.49%98.76%

Ave. ± Std99.86 ± 0.05%98.68 ± 0.52%99.48 ± 0.21%97.60 ± 0.83%99.50 ± 0.27%98.81 ± 0.42%

Note: gamma pass rate counted all the points with a dose >10% of the prescribed dose.