Research Article

Accuracy and Repeatability of the Gait Analysis by the WalkinSense System

Table 3

Percentage difference, intraclass correlation coefficient (ICC), and Pearson correlation coefficient (CC) between Pedar and WalkinSense during gait.

Percentage differences (%)ICCPearson CC
Average differenceCI95%ICCCI95%

−6.2−6.8−5.60.9730.9710.9750.869<0.001
3.43.13.60.9970.9970.9970.973<0.001
−14.1−14.9−13.20.9550.9520.9580.872<0.001
−13.2−14.1−12.40.9560.9530.9590.872<0.001

: peak pressure; : pressure time; : mean pressure; : pressure-time integral.