Rhythm Analysis during Cardiopulmonary Resuscitation: Past, Present, and Future

Figure 4

Filtering example for a VF recorded during OHCA. In the top panel the ECG is corrupted by CPR artifacts; a SAA from a commercial AED analyzes the rhythm every 10 s and gives erroneous no-shock (NS) diagnoses. In the bottom panel the CPR artifact is suppressed using an adaptive filter, the underlying VF is revealed, and the SAA gives correct shock (S) diagnoses. The underlying VF is visible in the 5 s interval without chest compressions.
386010.fig.004