Research Article

Semiautomatic Cochleostomy Target and Insertion Trajectory Planning for Minimally Invasive Cochlear Implantation

Table 2

DCA target accuracy and insertion results.

No.Target accuracy (mm)Insertion time (min)Intracochlear contactsAngular insertion depth (°)Implanted scala

1L0.79*58 of 12270SV
1R0.60*510 of 12330ST
2L0.0757 of 12210ST
2R0.1528 of 12300ST
3L0.05512 of 12420ST
3R0.33*211 of 12360ST
4L0.2858 of 12290ST
4R0.2479 of 12360ST
5L0.27*511 of 12300ST
5R0.22412 of 12350ST

Avg. ± SD0.30 ± 0.235 ± 210 ± 2319 ± 58

SV: scala vestibule; ST: scala tympani; *cases with broken screws which caused varying degrees of error in the registration.