Reproducibility of the Optical Biometer OA-1000 (Tomey)
Figure 3
Boxplots show for all 5 measurements in the measurement sequence the median, 25% and 75% quartile, as well as the 95% confidence interval of measurements. None of the measurement parameters show a drift from the first to last measurement.
(a) Measurement of axial length values (AL).
(b) Measurement of anterior chamber depth (ACD).
(c) Measurement of central corneal thickness (CT).