BioMed Research International
Volume 2014 (2014), Article ID 818694, 5 pages
Clinical Study

Comparison of Selected Kinematic Facebows Applied to Mandibular Tracing

1Division of Dental Materials, Faculty of Dentistry, Wroclaw Medical University, 50425 Wroclaw, Poland
2Department of Periodontology, Faculty of Dentistry, Wroclaw Medical University, 50425 Wroclaw, Poland
3Department of Prosthetic Dentistry, Faculty of Dentistry, Wroclaw Medical University, 50425 Wroclaw, Poland

Received 20 March 2014; Revised 21 April 2014; Accepted 22 April 2014; Published 7 May 2014

Academic Editor: Anna Paradowska-Stolarz

Copyright © 2014 Mieszko Wieckiewicz et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.


The study focused on the comparison between mechanical and computerized registration methods used by the two selected kinematic facebows. The material consisted of 35 women aged 18 to 35, studied using the Gerber Dynamic Facebow and the computerized ARCUSdigma II axiograph. To compare the devices the condylar path inclination (CPI) was recorded according to the Camper’s line, enabling the acquisition of easily comparable values based on which the devices were objectively and subjectively analyzed. Statistics was performed for the obtained data. The study showed that the values for the CPI registrated by the ARCUSdigma II are significantly higher than those obtained by using the Gerber Dynamic Facebow. The significant difference in the records of the CPI is most likely a result of the differences in the registration techniques assumptions. ARCUSdigma II provides the user with more diagnostic options than Gerber Dynamic Facebow. Mechanical facebow handling has a higher risk of hand-measuring errors in tracing procedure. Due to high discrepancy of achieved results from different systems the authors recommend to use articulator compatible with facebow whose measurement has been done.