Research Article
Wear Behavior of an Unstable Knee: Stabilization via Implant Design?
Table 1
Results of wear particle analysis.
| | Unstable knee | Stable knee | Effect size |
| Particles analysed | 2016 | 1510 | | Estimated number of particles per 106 cycles | 1.09 1012 ± 0.14 1012 | 0.80 1012 ± 0.12 1012 | 2.23 | Equivalent circle diameter | 0.263 ± 0.160 µm | 0.246 ± 0.162 µm | 0.11 | Aspect ratio | 1.776 ± 0.584 | 1.700 ± 0.504 | 0.14 | Roundness | 0.548 ± 0.151 | 0.577 ± 0.143 | 0.20 | Form factor | 0.657 ± 0.137 | 0.687 ± 0.120 | 0.23 |
|
|