Dose Distributions of an 192Ir Brachytherapy Source in Different Media
Table 1
Radial dose function,
(
), calculated by MCNPX code for the
microSelectron source centered in a spherical tissue phantom with
cm.
| Radial distance (cm) | Water | Bone | Lung | Radial distance (cm) | Water | Bone | Lung |
| 0.2 | 0.999 | 0.978 | 1.003 | 4.8 | 0.986 | 0.989 | 0.994 | 0.4 | 0.997 | 0.982 | 1.000 | 5.0 | 0.988 | 0.983 | 0.994 | 0.6 | 0.998 | 0.989 | 1.000 | 5.5 | 0.979 | 0.956 | 0.994 | 0.8 | 1.000 | 0.995 | 1.000 | 6.0 | 0.974 | 0.937 | 0.993 | 1.0 | 1.000 | 1.000 | 1.000 | 6.5 | 0.966 | 0.914 | 0.989 | 1.2 | 1.001 | 1.006 | 1.001 | 7.0 | 0.955 | 0.887 | 0.988 | 1.4 | 1.000 | 1.006 | 1.000 | 7.5 | 0.947 | 0.863 | 0.984 | 1.6 | 1.003 | 1.015 | 1.001 | 8.0 | 0.936 | 0.830 | 0.982 | 1.8 | 1.004 | 1.016 | 1.002 | 8.5 | 0.923 | 0.799 | 0.981 | 2.0 | 1.006 | 1.018 | 1.001 | 9.0 | 0.913 | 0.767 | 0.974 | 2.2 | 1.001 | 1.023 | 1.001 | 9.5 | 0.891 | 0.736 | 0.966 | 2.4 | 0.997 | 1.019 | 1.000 | 10.0 | 0.872 | 0.711 | 0.963 | 2.6 | 1.003 | 1.022 | 1.001 | 10.5 | 0.859 | 0.672 | 0.959 | 2.8 | 1.004 | 1.022 | 1.001 | 11.0 | 0.833 | 0.636 | 0.956 | 3.0 | 1.001 | 1.019 | 1.000 | 11.5 | 0.820 | 0.607 | 0.950 | 3.2 | 0.999 | 1.017 | 0.999 | 12.0 | 0.797 | 0.574 | 0.947 | 3.4 | 0.997 | 1.015 | 0.997 | 12.5 | 0.774 | 0.536 | 0.940 | 3.6 | 0.997 | 1.011 | 0.997 | 13.0 | 0.746 | 0.501 | 0.934 | 3.8 | 0.998 | 1.004 | 0.997 | 13.5 | 0.719 | 0.467 | 0.926 | 4.0 | 0.996 | 1.002 | 0.997 | 14.0 | 0.687 | 0.421 | 0.918 | 4.2 | 0.993 | 1.000 | 0.996 | 14.5 | 0.651 | 0.374 | 0.909 | 4.4 | 0.991 | 0.995 | 0.996 | 15.0 | 0.612 | 0.329 | 0.900 | 4.6 | 0.988 | 0.991 | 0.996 | | | | |
|
|