Saliva-Based Biosensors: Noninvasive Monitoring Tool for Clinical Diagnostics Table 1 Summary of the analytical parameters for catalytic salivary sensors reported in the literature.
Biomarker Biorecognition Transducer Sample volume ( L) Working range Sensitivity Detection limit (DL) Reference Glucose GOx Electrochemical 200 0.1–10 mg dL−1 — — [31 ] GOx Electrochemical 50 0.1–10 mg dL−1 — — [32 ] GOx/Fc* Electrochemical — 0–2.2 mM 21.45 nA mol−1 cm−2 1 M [33 ] GOx/POx/TMBZ* Optical 0.2 0–10 mg dL−1 — — [34 , 35 ] GOx/POx/TMBZ* Optical 0.2 — — — [19 ] Lactate LOx Electrochemical 1000 — 42 nA mM−1 — [4 , 42 ] LOx Electrochemical — 0.01–5 mM 78 nA mM−1 — [38 ] LOx/luminol* Electrochemiluminescent 50 0.1–0.5 mM — 5 M [36 ] LOx/PB* Electrochemical — 0.1–1.0 mM 0.553 A mM−1 50 M [44 ] LOx/PB* Electrochemical 4 0.1–5 mM 0.3169 A mM−1 300 M [45 ] Phosphate PyOx Electrochemical 50 7.5–625 M 523.89 nA mM−1 3.6 M [53 ] Alpha-amylase GD/GOx Electrochemical 100 0–30 U mL−1 — — [62 ] GD/MR/GOx Electrochemical 35 0.1–3 mM — 50/12 nM (5/30 min) [63 ] Gal-G2-CNP Optical 5 0–200 U mL−1 — — [66 ] Gal-G2-CNP Optical 30 10–140 U mL−1 — — [65 ] Gal-G2-CNP Optical 25 10–230 U mL−1 — — [64 ] MP/GOx/POx Electrochemical 50 0–190 U mL−1 — — [67 ] Amylase Ag-Ab* Immunoelectrochemical — 3 × 10−3 –1.6 × 10−2 ng mL−1 — 1.57 pg mL−1 [68 ] GD/GOx/Fc* Electrochemical — 60–840 U mL−1 — 17 U mL−1 [70 ] GD/MR/GOx/PB* Electrochemical — 5–250 U mL−1 14 nA U−1 5 U mL−1 [71 ]
The biorecognition molecule is listed together with the chromogen that provides the change of colour for optical determinations with the redox mediators that facilitate electrons transfer for electrochemical determinations.