Comparative Myocardial Deformation in 3 Myocardial Layers in Mice by Speckle Tracking Echocardiography

Table 1

Comparison between GE Vivid 7 and Vevo 2100 in 2D-speckle tracking hardware and software abilities.

Hardware abilities
GE Vivid 7Vevo 2100

Transducer frequency10–14 MHz18–38 MHz
Axial resolution200 µm50 µm
Lateral resolution300 µm110 µm
Transducer footprint28 mm × 10 mm20 mm × 5 mm
Temporal resolution130–190 fps>200 fps
Detect respiratory cycle YesYes
Image acquisition Conventional Hand-free + conventional
Area of analysisMyocardiumEndocardium + epicardium
ROI adjustmentUniformIndividual
Ability to distinguish E/A wave when heart rate is more than 400 bpmNoYes
Marked AVCYesNo
FAC ManualAutomated

fps: frame per second, AVC: aortic valve closure, and FAC: fractional area change (defined as cross-sectional area change between end diastole and end systole).