Research Article

Design and Evaluation of a Fully Implantable Control Unit for Blood Pumps

Table 1

Influence of the different requirements on each part of the control unit.

Input powerDriver electronicsMicrocontroller

Wireless 
safety
Add rectifier 
Add battery
Adapted to wide voltage input rangeMonitor TET voltage 
Monitor battery voltage

AutonomousAutomatic switching, recharging as soon as TET is connectedProvide coil currentsControl of the motor trajectory by control of the motor currents

EfficiencyUtilization of small number of efficient components with low heat production

GeometryUtilization of small packages and low number of electrical components