Review Article

Characterization of Red Blood Cells with Multiwavelength Transmission Spectroscopy

Table 3

The slopes and correlation coefficients of the regressions between the MWT analysis and SYSMEX SF-3000 values, bias, and standard deviation (S.D.) for the RBC indices (MCV, MCHC, MCH, RBC#, HCT, and HGB) calculated for the full data set of samples from anemic patients ( = 51) and for the reduced set that excluded outliers ( = 43).

Stat/parameterMCVMCHCMCHRBC#HCTHGB

Slope ( = 51)1.001.001.001.081.071.07
( = 51)0.720.700.700.680.610.65
Bias ( = 51)2.1%1.8%2.5%10%11%11%
S.D. ( = 51)5.00.81.40.32.80.8

Slope ( = 43)1.001.001.001.051.041.05
( = 43)0.990.930.980.860.860.85
Bias ( = 43)0.8%1.0%1.3%4.9%5.0%4.4%
S.D. ( = 43)0.60.30.250.21.40.5