Review Article

Theranostic Imaging of Yttrium-90

Table 1

Image acquisition parameters used for clinical 90Y bremsstrahlung planar and SPECT imaging studies.

ReferenceImaging90Y agentCollimatorEnergy window(s)
keV
Attenuation
correction

Smith et al. [6]PlanarSilicateMedium energy60–200No
Tehranipour et al. [27]PlanarResin microspheresMedium energy72–119No
Minarik et al. [26]PlanarAnti-CD20 antibodyHigh energy105–195No
Ahmadzadehfar et al. [31]PlanarResin microspheresMedium energy55–250No
Ahmadzadehfar et al. [30]PlanarResin microspheresMedium energy55–250No
Smith et al. [6]SPECTSilicateMedium energy60–200No
Mansberg et al. [48]SPECTResin microspheresMedium energy77–104Yes
Flamen et al. [49]SPECTResin microspheresMedium energy53–88 and 97–287Yes
Minarik et al. [2]SPECTAnti-CD20 antibodyHigh energy105–195Yes
Lhommel et al. [50]SPECTResin microspheresMedium energy77–104Yes
Minarik et al. [26]SPECTAnti-CD20 antibodyHigh energy105–195Yes
Strigari et al. [29]SPECTResin microspheresMedium energy55–245Yes
Ahmadzadehfar et al. [31]SPECTResin microspheresMedium energy55–250Yes
Ahmadzadehfar et al. [30]SPECTResin microspheresMedium energy55–250Yes
Wissmeyer et al. [62]SPECTGlass microspheresMedium energy77–104Yes
Fabbri et al. [47]SPECTDOTATOCMedium energy58–102 and 153–187Yes
Elschot et al. [55]SPECTResin microspheresHigh energy50–250Yes
Elschot et al. [45]SPECTResin microspheresHigh energy105–195Yes
Kao et al. [14, 58]SPECTResin microspheresMedium energy74–86Yes
Padia et al. [61]SPECTGlass microspheresMedium energy57–100Yes
Ulrich et al. [56]SPECTResin microspheresMedium energy68–83Yes
Wondergem et al. [57]SPECTResin microspheresHigh energy50–250Yes
Eaton et al. [59]SPECTResin microspheresMedium energy55–95Yes