Research Article
Performances and Reliability of Bruker Microflex LT and VITEK MS MALDI-TOF Mass Spectrometry Systems for the Identification of Clinical Microorganisms
Table 2
Results of 18 reference bacterial strains that were analyzed by VITEK MS and Bruker Microflex LT for an initial system setup and comparison.
| Microorganism | VITEK | Bruker |
| S. pyogenes ATCC 19615 | ✓ | ✓ |
| E. coli ATCC 25922 | ✓ | ✓ |
| K. pneumoniae ATCC 13883 | ✓ | ✓ |
| P. aeruginosa ATCC 27853 | ✓ | ✓ |
| S. aureus ATCC 25923 | ✓ | ✓ |
| S. pneumoniae ATCC 49619 | ✓ | ✓ |
| P. vulgaris ATCC 13315 | P. vulgaris/penneri | P. hauseri (new nomenclature) |
| S. typhimurium ATCC 14028 | Salmonella group | Salmonella spp. |
| Y. enterocolitica ATCC 9610 | ✓ | ✓ |
| V. cholera RSKK 913 | ✓ | ✓ (V. cholerae biovar albensis) |
| G. vaginalis DM 108 | ✓ | ✓ |
| S. saprophyticus DM 17 | ✓ | ✓ |
| S. agalactiae DM 20 | ✓ | ✓ |
| MRSA DM 66 | S. aureus | S. aureus |
| S. epidermidis DM 19 | ✓ | ✓ |
| S. enteritidis DM 10 | Salmonella group | Salmonella spp. |
| A. hydrophila DM 110 | A. hydrophila/caviae | A. caviae |
| P. shigelloides DM 111 | ✓ | ✓ |
|
|