Research Article

Performances and Reliability of Bruker Microflex LT and VITEK MS MALDI-TOF Mass Spectrometry Systems for the Identification of Clinical Microorganisms

Table 5

Results summary of MALDI-TOF MS systems for the microorganisms that either were misidentified or could not be identified.

Microorganism groupsNumber of isolatesSpecies contentMisidentificationNo-identification
VITEKBrukerVITEKBruker

Nonfermentatives
Acinetobacter baumannii1942 (F × 2)2 (F × 2)
Aeromonas sobria11 (F)
Total35616223

Enterobacteriaceae
Kluyvera cryocrescens11 (G)1 (G)
Shigella boydii11 (N/A)1 (N/A)
Total309231111

Staphylococci
Staphylococcus epidermidis682 (C × 2)1 (F)
Total2731021

Streptococci
Leuconostoc mesenteroides11 (F)1 (F)
Total291111

Enterococci and other Gr(+) cocci
Total1605

Coryneforms and other Gr(+) bacilli
Actinomyces naeslundii11 (F)
Corynebacterium coyleae11 (D)
Corynebacterium glaucum11 (G)1 (G)
Corynebacterium mucifaciens31 (D)2 (F)
Corynebacterium ureicelerivorans21 (F)
Total3014251

HACEK and other Gr(−) cocci
Total188

Anaerobes
Total22

Others
Wautersiella falsenii11 (G)1 (G)
Total441 (G)1 (G)

Total (bacteria)1,1819354107

Fungi
Aspergillus brasiliensis11 (F)
Total (fungi)160141

Total (bacteria + fungi)1,34110754117