Research Article

Changes of the eSheath Outer Dimensions Used for Transfemoral Transcatheter Aortic Valve Replacement

Table 1

Summary of unexpanded, maximum, and mean eSheath outer diameters of 14 F and 16 F eSheaths.

eSheath outer diameter (mm)14 F eSheath/23 mm Sapien 14 F eSheath/26 mm Sapien 16 F eSheath/29 mm Sapien value
(23 versus 29 mm)

Unexpanded eSheath 5.80 ± 0.12 mm 5.80 ± 0.12 mm 6.50 ± 0.11 mm 0.0001
eSheath ex vivo  max. at pusher site7.65 ± 0.34 mm7.64 ± 0.42 mm 8.18 ± 0.27 mm 0.03
Mean eSheath diameter after TAVI7.14 ± 0.45 mm 7.26 ± 0.40 mm 8.10 ± 0.58 mm 0.02