Research Article
Changes of the eSheath Outer Dimensions Used for Transfemoral Transcatheter Aortic Valve Replacement
Table 1
Summary of unexpanded, maximum, and mean eSheath outer diameters of 14 F and 16 F eSheaths.
| eSheath outer diameter (mm) | 14 F eSheath/23 mm Sapien | 14 F eSheath/26 mm Sapien | 16 F eSheath/29 mm Sapien | value (23 versus 29 mm) |
| Unexpanded eSheath | 5.80 ± 0.12 mm | 5.80 ± 0.12 mm | 6.50 ± 0.11 mm | 0.0001 | eSheath ex vivo max. at pusher site | 7.65 ± 0.34 mm | 7.64 ± 0.42 mm | 8.18 ± 0.27 mm | 0.03 | Mean eSheath diameter after TAVI | 7.14 ± 0.45 mm | 7.26 ± 0.40 mm | 8.10 ± 0.58 mm | 0.02 |
|
|