Validity of Four Commercial Bioelectrical Impedance Scales in Measuring Body Fat among Chinese Children and Adolescents

Table 3

Cross-validation of four BIA devices (embedded equations) for the perdition of %BF/FFM in Chinese children and adolescents.

GenderBIA devices%BF-DXA%BF-BIAFFM-DXAFFM-BIA  -testBias (SD)TE95%
limits of agreement

Boys ()Model A20.77%21.34%0.2680.58 (5.86)%5.86%0.513−10.96% to 12.12%
Model B20.77%21.92%0.0311.15 (5.95)%6.04%0.528−10.51% to 12.81%
Model C20.77%19.04%<0.001−1.67 (4.80)%5.06%0.162−11.08% to 7.74%
Model D20.77%20.22%0.246−0.37 (3.51)%3.51%0.365−7.25% to 6.51%
Model A43.02 kg43.80 kg0.0190.78 (3.70) kg3.77 kg0.261−6.47 kg to 8.03 kg
Model B43.02 kg43.18 kg0.6080.16 (3.60) kg3.60 kg−0.027−6.90 kg to 7.22 kg
Model C43.02 kg44.50 kg<0.0011.54 (3.35) kg3.60 kg0.024−5.03 kg to 8.11 kg
Model D43.02 kg44.31 kg<0.0011.02 (2.17) kg2.39 kg0.197−3.23 kg to 5.27 kg

Girls ()Model A27.85%27.37%0.266−0.48 (4.88)%4.88%0.359−10.04% to 9.08%
Model B27.85%25.22%<0.001−2.63 (3.23)%4.15%−0.049−8.96% to 3.70%
Model C27.85%25.81%<0.001−2.05 (3.21)%3.79%0.012−8.34% to 4.24%
Model D27.85%26.34%<0.001−1.47 (2.99)%3.31%−0.147−7.33% to 4.39%
Model A32.77 kg33.89 kg<0.0011.11 (2.42) kg2.66 kg0.245−3.63 kg to 5.85 kg
Model B32.77 kg34.74 kg<0.0011.97 (1.77) kg2.64 kg−0.255−1.50 kg to 5.44 kg
Model C32.77 kg34.49 kg<0.0011.72 (1.77) kg2.47 kg−0.219−1.75 kg to 5.19 kg
Model D32.77 kg34.03 kg<0.0011.25 (1.22) kg1.68 kg−0.022−1.14 kg to 3.64 kg

Note: %BF: body fat percentage; FFM: fat-free mass; DXA: dual-energy X-ray absorptiometry; BIA: bioelectrical impedance analysis.
Bias: value obtained from BIA measurement minus DXA measurement; SD: standard deviation.
, Pearson’s correlation coefficient for the relationship between averaged %BF/FFM [(value of DXA + value of BIA)/2] and the bias.
; .
Model A: Biodynamics-310; Model B: Tanita TBF 543; Model C: Tanita BC-545; Model D: Inbody 520.