Patient-Specific Simulation of Coronary Artery Pressure Measurements: An In Vivo Three-Dimensional Validation Study in Humans

Table 2

Comparison between the wire-based measured pressure values () and the computed values from the four types of simulations ().

Patient #
(mm Hg)
FSI-transient
(mm Hg)
Rigid-transient
(mm Hg)
FSI-steady state
(mm Hg)
Rigid-steady state
(mm Hg)

1-RCA103.04105.93106.06105.46105.48
2-RCA104.73103.52103.38103.82103.71
3-LAD89.3289.038989.1789.15
4-RCA61.9563.6863.8364.0164.22
4-LAD67.4770.7271.4570.9471.59
5-RCA105.2102.42102.68102.51102.58
5-LAD85.9583.6483.8583.5983.72
6-LAD83.5282.882.7782.9883.03