BioMed Research International
Volume 2015, Article ID 684250, 10 pages
http://dx.doi.org/10.1155/2015/684250
Clinical Study

Volumetric MR-Guided High-Intensity Focused Ultrasound with Direct Skin Cooling for the Treatment of Symptomatic Uterine Fibroids: Proof-of-Concept Study

1Department of Radiology, University Medical Center Utrecht, Heidelberglaan 100, 3584 CX Utrecht, The Netherlands
2Philips Healthcare, Philips Medical Systems MR, Äyritie 4, 01511 Vantaa, Finland
3Department of Radiology and Image Sciences Institute, University Medical Center Utrecht, Heidelberglaan 100, 3584 CX Utrecht, The Netherlands

Received 11 July 2014; Accepted 18 August 2014

Academic Editor: Trevor Andrews

Copyright © 2015 Marlijne E. Ikink et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

Objective. To prospectively assess the safety and technical feasibility of volumetric magnetic resonance-guided high-intensity focused ultrasound (MR-HIFU) ablation with direct skin cooling (DISC) during treatment of uterine fibroids. Methods. In this proof-of-concept study, eight patients were consecutively selected for clinical MR-HIFU ablation of uterine fibroids with the use of an additional DISC device to maintain a constant temperature () at the interface between the HIFU table top and the skin. Technical feasibility was verified by successful completion of MR-HIFU ablation. Contrast-enhanced T1-weighted MRI was used to measure the treatment effect (nonperfused volume (NPV) ratio). Safety was evaluated by recording of adverse events (AEs) within 30 days’ follow-up. Results. All MR-HIFU treatments were successfully completed in an outpatient setting. The median NPV ratio was 0.56 (IQR [0.27–0.72]). Immediately after treatment, two patients experienced coldness related discomfort which resolved at the same day. No serious (device-related) AEs were reported. Specifically, no skin burns, cold injuries, or subcutaneous edema were observed. Conclusion. This study showed that it is safe and technically feasible to complete a volumetric MR-HIFU ablation with DISC. This technique may reduce the risk of thermal injury to the abdominal wall during MR-HIFU ablation of uterine fibroids. This trial is registered with NTR4189.