Auditing of Monitoring and Respiratory Support Equipment in a Level III-C Neonatal Intensive Care Unit

Table 5

Rates for appropriate overall use and variable with lower rates of appropriate programming of the different pieces of equipment in relation to time characteristics. SpO2: oxygen saturation; MAP: mean airway pressure; Vol/min alarms: volume/minute alarms; CPAP: continuous positive airway pressure.

Technological resources ( assessments)DayShiftQuarter
Normal working dayNonworking dayEarlyLate3rd quarterRest of year

Pulse oximeter ( 94)
 % appropriate overall use ()64.56 (51) 46.67 (7)0.19 65.12 (28) 58.82 (30) 0.53 76.92 (20) 55.88 (38) 0.06
Multimeasurement monitor ( 104)
 % appropriate overall use ()16.05 (13) 39.13 (9)0.01 16.22 (12) 33.33 (10) 0.05 19.35 (6)21.92 (16) 0.76
Conventional mechanical ventilator ( 53)
 % appropriate overall use ()8.51 (4)33.33 (2)0.07 10.34 (3)12.50 (3)0.80 18.75 (3)8.11 (3)0.26
High-frequency oscillatory ventilator ( 64)
 % appropriate overall use ()25.53 (12) 70.59 (12) 0.001 38.46 (15) 36 (9)0.84 35 (7)38.64 (17) 0.78
CPAP ( 68)
 % appropriate overall use ()27.91 (12) 28 (7)0.99 22.86 (8)33.33 (11) 0.33 35.71 (5)25.93 (14) 0.46
Technology as a whole ( 383)
 % appropriate overall use ()30.98 (92) 43.02 (37) 0.03 30 (66) 38.65 (63) 0.07 38.32 (41) 31.88 (88) 0.23