Research Article

A Randomized Comparison Simulating Face to Face Endotracheal Intubation of Pentax Airway Scope, C-MAC Video Laryngoscope, Glidescope Video Laryngoscope, and Macintosh Laryngoscope

Table 2

Comparison of intubation time (sec) and POGO (%) according to laryngoscopes (mean ± SD).

MCLAWSGVLC-MAC value

VCET (sec)7.8 ± 3.310.9 ± 7.88.4 ± 4.98.4 ± 4.60.199
POGO (%)53.6 ± 22.381.7 ± 18.365.4 ± 25.072.9 ± 20.80.000
Tube pass time (sec)18.7 ± 7.319.6 ± 9.526.8 ± 10.022.8 ± 10.20.001
1st ventilation time (sec)28.4 ± 7.729.6 ± 10.939.2 ± 9.735.2 ± 10.40.000
VCET to tube pass time (sec)10.9 ± 5.910.4 ± 10.922.8 ± 27.117.1 ± 14.30.000
Tube pass time to 1st ventilation time (sec)9.6 ± 3.99.8 ± 3.710.6 ± 9.712.2 ± 4.50.000

value < 0.05 is level of statistical significance according to Friedman test.
MCL: Macintosh laryngoscope; AWS: Pentax airway scope; GVL: Glidescope video laryngoscope; C-MAC: C-MAC video laryngoscope.