Research Article

The Effect of Femoral Cutting Guide Design Improvements for Patient-Specific Instruments

Table 3

Differences between planned bone resections and bone resections recorded from the PSI intraoperatively (mm).

Group IIGroup III
MRI 2 mm slice thickness (w/motion artifacts) & offset design ()MRI 2 mm slice thickness (w/o motion artifacts) & full contact design ()Ave.MRI 2 mm slice thickness (w/motion artifacts), offset & optimized design ()MRI 2 mm slice thickness (w/o motion artifacts), full contact & optimized design ()Ave.

Distal (±SD)0.56 (±0.12)0.28 (±0.04)0.44 (±0.17)0.26 (±0.04)0.16 (±0.04)0.21 (±0.06)
(±SD)0.46 (±0.09)0.18 (±0.03)0.32 (±0.15)0.14 (±0.03)0.12 (±0.02)0.13 (±0.03)

Posterior (±SD)0.34 (±0.07)0.23 (±0.03)0.27 (±0.08)0.20 (±0.03)0.18 (±0.02)0.19 (±0.02)
(±SD)0.19 (±0.03)0.17 (±0.01)0.18 (±0.02)0.14 (±0.02)0.12 (±0.01)0.13 (±0.01)