Assessment of Heat Hazard during the Polymerization of Selected Light-Sensitive Dental Materials

Table 1

Types of materials, sample sizes, sample maximum temperature , and time of reaching maximum temperature (: mass, : diameter, : height, and : the distance between the polymerization unit and the sample).

Material (mg) (mm) (mm) (mm) (°C) (s)

Te-Econom A-17024040.57.6
7024339.28.2
15054039.817.5
22028043.510.6
52058052.520.6

Filtek Supreme XT A-1B12054039.217.0
49058041.323.4

Filtek Supreme XT A-3B21928046.822.5

Filtek Supreme XT A-1D6924031.49.3
7124036.743.0
12354037.149.8
20028041.014.0
20028340.814.4

Filtek Supreme XT A-4D20028344.249.0

Filtek Supreme XT A-1E6824336.410.0
20028342.214.2

Tetric EvoCeram6924031.122.9
7024334.211.4
14054038.521.8
22028042.110.0
21028339.343.1
50058044.840.4

Gradia Direct A-36024036.844.6
6024336.149.6
9254051.641.4
17028055.541.9
17228362.542.0
38058064.339.7

Ionosit180.12.7034.313.3
190.13.0235.412.9

ExciTE220.16.0030.23.1
210.15.0242.24.3
220.15.5539.44.9

E: enamel, D: dentine, and B: body.