Research Article

Accuracy and Precision of Three-Dimensional Low Dose CT Compared to Standard RSA in Acetabular Cups: An Experimental Study

Table 2

Accuracy for 3D CT.

2.57 RMSMeanMinMax

Uncemented cup
 Translation (mm)
  0.290.05−0.110.19
  0.28−0.06−0.130.05
  0.320.04−0.060.26
 Rotation (d)
  0.82−0.15−0.560.17
  0.71−0.18−0.520.06
  0.43−0.04−0.220.10
Cemented cup
 Translation (mm)
  0.190.01−0.170.17
  0.070.01−0.080.07
  0.080.00−0.100.07
 Rotation (d)
  0.210.01−0.150.28
  0.440.05−0.430.49
  0.26−0.01−0.330.22