Research Article
Factors Associated with Missed Detection of Mycobacterium tuberculosis by Automated BACTEC MGIT 960 System
Table 1
Proportion of false-negative culture results determined by MGIT among specimens belonging to different grades of AFB smear.
| Grade of AFB smear | Total | Number of positive cultures (%) | MGIT positiv | MGIT false negative |
| Negative | 494 | 59 (11.9) | 4 (0.8) | Scanty | 11 | 9 (81.8) | 1 (9.1) | 1+ | 23 | 20 (87.0) | 1 (4.3) | 2+ | 22 | 21 (95.5) | 1 (4.5) | 3+ | 18 | 17 (94.4) | 0 (0.0) | 4+ | 9 | 7 (77.8) | 1 (11.1) | Total | 577 | 133 (23.1) | 8 (1.4) |
|
|
: BACTEC MGIT 960; , .
|