The Effect of File Size and Type and Irrigation Solutions on the Accuracy of Electronic Apex Locators: An In Vitro Study on Canine Teeth

Table 1

Basic statistics of the measurement results of the distance between the instrument and the apical constriction performed in the 5.25% NaOCl environment with the use of steel S-files (Poldent) with the size ranging from 08 to 30 in accordance with ISO and with the use of two apex locators: Endopilot and iPex NSK as well as the results of comparison made with the use of Student’s -test for dependent variables.

Instrument sizeEndopilotiPex NSK

 mm, ⌀8 mm3.23 ± 0.032.58 ± 0.0327.577<0.001
 mm, ⌀10 mm2.93 ± 0.122.22 ± 0.107.9850.001
 mm, ⌀15 mm3.02 ± 0.122.07 ± 0.158.5930.001
 mm, ⌀20 mm2.58 ± 0.102.00 ± 0.058.7500.001
 mm, ⌀25 mm2.48 ± 0.061.98 ± 0.0313.416<0.001
 mm, ⌀30 mm2.08 ± 0.061.88 ± 0.132.3700.077
 mm, ⌀8 mm1.82 ± 0.081.70 ± 0.101.6060.184
 mm, ⌀10 mm1.73 ± 0.151.67 ± 0.030.4990.499
 mm, ⌀15 mm1.52 ± 0.061.38 ± 0.033.5780.023
 mm, ⌀20 mm1.37 ± 0.031.30 ± 0.052.0000.116
 mm, ⌀25 mm1.25 ± 0.001.23 ± 0.031.0000.374
 mm, ⌀30 mm1.12 ± 0.081.22 ± 0.03−2.1210.101
 mm, ⌀8 mm1.48 ± 0.081.38 ± 0.032.1210.101
 mm, ⌀10 mm1.38 ± 0.081.32 ± 0.031.4140.230
 mm, ⌀15 mm1.15 ± 0.051.03 ± 0.033.5000.025
 mm, ⌀20 mm0.93 ± 0.030.83 ± 0.034.2430.013
 mm, ⌀25 mm0.78 ± 0.030.62 ± 0.037.0710.002