Research Article

The Effect of File Size and Type and Irrigation Solutions on the Accuracy of Electronic Apex Locators: An In Vitro Study on Canine Teeth

Table 2

Basic statistics of the distance between the endodontic device and the apical constriction measured with the use of a steel file with the size of 30/22 mm for different irrigation solutions.

Irrigation solutionEndopilotiPex NSK

2.0% NaOCl0.51 ± 0.020.49 ± 0.021.6040.184
5.25% NaOCl0.50 ± 0.050.52 ± 0.01−0.6110.574
40.0% citric acid2.10 ± 0.041.51 ± 0.0811.693<0.001
2.0% Gluco-chex2.11 ± 0.051.99 ± 0.062.6680.056
Isopropyl alcohol00
3.0% H2O22.13 ± 0.032.02 ± 0.082.2200.091
15.0% Endosal1.37 ± 0.101.49 ± 0.06−1.8860.132
0.9% NaCl2.07 ± 0.031.23 ± 0.0524.362<0.001

Statistically significant differences between the measurement results obtained with the use of the Endopilot and iPex NSK apex locators were observed in the 40.0% citric acid and 0.9% NaCl solutions ().