Backside Wear Analysis of Retrieved Acetabular Liners with a Press-Fit Locking Mechanism in Comparison to Wear Simulation In Vitro

Table 3

Average score per wear mode of all backside sections from in vitro and retrieved liners (maximum possible score per wear mode = 3).

Wear modeIn vitroRetrievals

Pitting0.00 (±1.04)0.03 (±0.17)
Scratching1.62 (±1.62)1.36 (±0.83)
Burnishing0.23 (±0.47)0.13 (±0.41)
Abrasion0.67 (±0.95)0.32 (±0.65)
Embedded particles0.07 (±0.24)0.70 (±0.82)
Deformation0.01 (±0.11)0.03 (±0.17)
Delamination0.00 (±0.00)0.00 (±0.00)