Backside Wear Analysis of Retrieved Acetabular Liners with a Press-Fit Locking Mechanism in Comparison to Wear Simulation In Vitro
Table 3
Average score per wear mode of all backside sections from in vitro and retrieved liners (maximum possible score per wear mode = 3).
| Wear mode | In vitro | Retrievals |
| Pitting | 0.00 (±1.04) | 0.03 (±0.17) | Scratching | 1.62 (±1.62) | 1.36 (±0.83) | Burnishing | 0.23 (±0.47) | 0.13 (±0.41) | Abrasion | 0.67 (±0.95) | 0.32 (±0.65) | Embedded particles | 0.07 (±0.24) | 0.70 (±0.82) | Deformation | 0.01 (±0.11) | 0.03 (±0.17) | Delamination | 0.00 (±0.00) | 0.00 (±0.00) |
|
|