Research Article
Comparison of the Feasibility and Safety of First- versus Second-Generation AMPLATZER™ Occluders for Left Atrial Appendage Closure
Table 5
Prevalence and severity of peridevice leaks at transesophageal echocardiographic follow-up.
| | ACP (%) () | Amulet (%) () | |
| Device-related thrombus | 4 (5) | 1 (1) | 0.17 | Device embolization | 0 (0) | 0 (0) | 1 | Peridevice leakage | 12 (15) | 4 (5) | 0.03 | Severe leaks | 0 (0) | 0 (0) | 1 | Major leak | 0 (0) | 0 (0) | 1 | Moderate leak | 0 (0) | 0 (0) | 1 | Minor leak | 12 (15) | 4 (5) | 0.03 |
|
|