| Scan parameters | Standard of care protocol | Protocol A | Protocol B |
| Tube voltage (kVp) | 120 | 80 | 100 | Tube current (mA) | 235 | 700 | 375 | (mGy) | 28.8 | 28.8 | 28.8 | Scan type | Cardiac helical | Cardiac helical | Cardiac helical | Scan delay (s) | 30 | 30 | 30 | Image reconstruction model | FBP | ASiR 60% | ASiR 60% | Contrast media concentration (mg I/mL) | 320 | 160 | 200 | Injection volume (mL) | 60 | 60 | 60 | Injection rate (mL/s) | 3 | 3 | 3 | Iodine delivery rate (g I/s) | 0.96 | 0.48 | 0.60 | Total iodine dose (mg I/kg) | 480 | 240 | 300 |
|
|
mg I/mL: milligrams of iodine per milliliter. g I/s: grams of iodine per second. mg I/kg: milligrams of iodine per kilogram of body weight.
|