Clinical Study

MR Lymphangiography: A Practical Guide to Perform It and a Brief Review of the Literature from a Technical Point of View

Table 1

Imaging parameters for magnetic resonance lymphangiography at 1,5 T.

TRTETIFA (°)FOV (cm)MatrixThickness/overlap (mm)NEXBandwidth (khz)

Coronal 3D SSFP balanced4.01.99040 × 40224 × 1922/10.53±125
Coronal 3D spoiled GRE T1W with SPECtral inversion at lipid balanced5.02.1172544 × 44448 × 3202.8/1.41±111.1
3D T2-weighted turbo spin-echo200068040 × 40320 × 2243.5/11±31.2

TR = repetition time; TE = echo time: TI = inversion time; FA = flip angle; FOV = field of view; NEX = number of excitations.
Sequences performed in our experience.